Datasheet

DC POWER SUPPLIES
www.keithley.com
1.888.KEITHLEY
(U.S. only)
A Greater Measure of Confidence
Series 2290
High Voltage Power Supplies
2290-10 Specifications
10kV Voltage Power Supply Characteristics
CONDITIONS
This document contains specifications and supplemental information for the Model 2290-10
High Voltage Power Supply. Specifications are the standards against which the Model 2290-10 is
tested. Upon leaving the factory, the Model 2290-10 meets these specifications. Characteristics,
supplemental characteristics, and typical values are not warranted, apply at 23°C ±5°C, <70%
relative humidity, and are provided solely as useful information.
CHARACTERISTICS
VOLTAGE RANGE:
Output Voltage
1
Maximum Output Current
+100 to +10,000 V DC 1.000 mA DC
VOLTAGE SET ACCURACY: ±0.06% of full scale.
VOLTAGE DISPLAY ACCURACY: Voltage set accuracy ±1V, typical (±2V, maximum).
VOLTAGE RESOLUTION: 1V (set and display).
VOLTAGE LIMIT RANGE: 0 to 100% full scale.
VOLTAGE REGULATION
2
: Line: 0.001% for ±10% line voltage change.
Load: 0.04% for 100% load change, typical.
OUTPUT RIPPLE (300Hz–300kHz): 0.01% of full scale, V
RMS
.
RISE TIME (from high voltage turn on to final value under full load)
3, 4
: <6 seconds to within
1V of the final value.
DISCHARGE TIME (to 0V under full load)
3, 4
: 1 second to under 10V.
DISCHARGE TIME (no load)
3
: <6 seconds (to <1% of full scale voltage with no load, typical).
SETTLING TIME
3, 4
: <4 seconds to within 1V of the final value.
RECOVERY TIME
3, 4
: 120ms for 40% step change in load current (typical).
CURRENT LIMIT RANGE: 0 to 105% of full scale.
CURRENT SET ACCURACY: 0.5% of full scale.
CURRENT RESOLUTION:A.
CURRENT DISPLAY ACCURACY: ±1µA, typical (±2µA, maximum).
TRIP CURRENT RANGE: 10µA to 105% of full scale (excluding stored output charge).
OUTPUT STORED CHARGE: <20µC maximum.
TRIP RESPONSE TIME: <10ms.
STABILITY: ±0.01% per hour; <0.03% per eight hours.
TEMPERATURE DRIFT: 50ppmC, 0º to 40ºC, typical.
PROTECTION: Arc and short circuit protected; programmable voltage and current limits and
current trip.
MONITOR OUTPUTS
OUTPUT SCALE: 0 to +10V for 0V to full scale.
CURRENT RATING: 10mA maximum.
OUTPUT IMPEDANCE: 100Ω.
ACCURACY: ±0.2% of full scale with a 1MΩ load, minimum.
UPDATE RATE: 87.5Hz.
EXTERNAL VOLTAGE SET
INPUT SCALE: 0 to +10V for 0 to full scale.
INPUT IMPEDANCE: 1MΩ.
ACCURACY: ±0.2% of full scale.
UPDATE RATE: 87.5Hz.
OUTPUT SLEW RATE (5% to 95% under full load): <3 seconds
GENERAL
INPUT POWER: 75 watts.
INPUT VOLTAGE: 90V–264VAC, 47 to 63Hz.
REAR PANEL CONNECTORS:
Output high-voltage connector: SHV male (Kings type 1764-1 or equivalent).
BNC Connector (three): Input, voltage monitor; current monitor.
GPIB connector, 23-pin.
RS-232, 9-pin D-sub
HIGH-VOLTAGE SAFETY INTERLOCK:
Connector: 3-pin press-in connector, 3M part number: 37103-A165-00E-MB.
Pin 1: 5V out, 70mA maximum out.
Pin 2: Input: High-voltage output enabled:5V DC.
High-voltage output disabled: 0V DC or open connection.
Pin 3: Chassis ground.
INTERFACE PROTOCOL: IEEE-488.1.
OPERATING ENVIRONMENT: 0º to 40ºC; non-condensing.
DIMENSIONS: 89mm high × 206mm wide × 406mm deep (3.5 in × 8.1 in × 16 in).
WEIGHT: 3.7kg (8 pounds).
SAFETY: Conformance to European Union low voltage directive.
WARR ANT Y: One year.
WARM-UP TIME: One hour.
1. The output voltage can be programmed to voltages below 100V, but accuracy below 100V is not specified.
2. Regulation specifications apply for greater than 100V.
3. Times apply with the current limit set to 105% of full scale.
4. Loads are fully resistive.
Model 2290-10 rear panel showing the IEEE-488 interface connector,
RS-232 interface connector, high voltage output connector, analog
control/output connectors, and interlock connector.
High voltage power supplies
High voltage power supplies