2290-PM-200 Specifications
SMU Protection Module Characteristics
CONDITIONS
This document contains specifications and supplemental information for the Model 2290-PM-
200 Protection Module. Specifications are the standards against which the Model 2290-PM-200
is tested. Upon leaving the factory, the Model 2290-PM-200 meets these specifications.
Characteristics, supplemental characteristics, and typical values are not warranted, apply at 23°C
±5°C, <70 percent relative humidity, and are provided solely as useful information.
CHARACTERISTICS
LEAKAGE: <10pA at 200V typical.
DC CURRENT CARRYING CAPABILITY: <1.5A in unclamped condition.
PULSED CURRENT CARRYING CAPABILITY: 10A maximum at 1% duty cycle in unclamped
condition.
MAXIMUM PROTECTION ACTIVE CURRENT: Current must be limited to <120mA in clamped
condition (~230V).
PROTECTION VOLTAGE: 220VDC minimum, 230VDC typical, 240VDC maximum.
DIMENSIONS: 81.28mm long × 81.53mm wide × 54.10mm high (3.20 in. long × 3.21 in. wide ×
2.13 in. high) Maximum length with high-voltage connectors: 5.34 in (135.64mm).
WEIGHT: 0.25kg (0.55 lb.).
GENERAL
WARR ANT Y: 1 year.
SAFETY: Listed to UL 61010-1:2012 (3rd edition). Conforms to European Union Low Voltage
Directive.
CONNECTORS:
High Voltage: 5kV connector: male SHV, 0.38 inches (9.35mm) diameter, Kings 1704-1.
10kV connector: female SHV, 0.535 inches (13.59mm) diameter, Kings 1764-1.
Low Voltage: Triax, female, quantity of 2
SAFETY: Screw ground terminal, quantity of 2