Datasheet
Datasheet
TEK.COM6
Test Script Specifications
2450-EC Cyclic Voltammetry Specifications
Potential Range –20 V ≤ E ≤ +20 V
Current Measure Ranges 10 μA, 100 μA, 1 mA, 10 mA, 100 mA, 1 A
Source Limit (Compliance) 100% of selected current range
Scan Rate 0.1 mV/s to 3500 mV/s
Potential Step Size During Scanning
100 μV (0.1 mV/s ≤ scan rate < 35 mV/s)
1 mV (35 mV/s ≤ scan rate < 350 mV/s )
10 mV (350 mV/s ≤ scan rate ≤ 3500 mV/s)
Number of Cycles 1 to 100
User Selectable Sampling Intervals Points/Test (10 to 10000)
Points/Cycle (10 to 10000)
Seconds/Point (0.01 to 100)
Points/Second (0.01 to 100)
Maximum Total Number of Samples 100,000
2450-EC Open Circuit Potential Specifications
Potential Measure Ranges 20 mV, 200 mV, 2 V, 20 V
Sample Interval 0.75 s ≤ sample interval ≤ 100 s
Number of Samples 1 to 100,000
2450-EC Potential Pulse and Square Wave Specifications
Peak Potential –20 V ≤ Epeak ≤ +20 V
Base Potential –20 V ≤ Ebase ≤ +20 V
Current Measure Ranges 1 μA, 10 μA, 100 μA, 1 mA, 10 mA, 100 mA, 1 A
Source Limit (Compliance) 105% of selected current range
Pulse Period and Width
Current Measure Range = 1 μA
200 ms ≤ period ≤ 3600 s
100 ms ≤ pulse width ≤ (0.99 × period)
Current Measure Range > 1 μA
4 ms ≤ period ≤ 3600 s
2ms ≤ pulse width ≤ (0.99 × period)
Sample Time Minimum is 166.667 μs.
Maximum is lesser of 166.667 ms and (pulse width – 1 ms)
Number of Cycles 1 to 100,000
Program Time (1 × period) ≤ program time ≤ (100,000 × period)