RESISTANCE MEASUREMENT ACCURACY (LOCAL OR REMOTE SENSE)
1,2,3
Range Default
resolution
6
Default test current Normal accuracy
23 °C ± 5 °C 1 year
± (% reading + ohms)
Enhanced accuracy
7
23 °C ± 5 °C 1 year
± (% reading + ohms)
< 2.000000 Ω
8
1 µΩ User-defined Source I
ACC
+ Meas V
ACC
Meas I
ACC
+ Meas V
ACC
20.00000 Ω 10 µΩ 100 mA 0.098 % + 0.003 0.073 % + 0.001
200.0000 Ω 100 µΩ 10 mA 0.077 % + 0.03 0.053 % + 0.01
2.000000 kΩ 1 mΩ 1 mA 0.066 % + 0.3 0.045 % + 0.1
20.00000 kΩ 10 mΩ 100 µA 0.063 % + 3 0.043 % + 1
200.0000 kΩ 100 mΩ 10 µA 0.065 % + 30 0.046 % + 10
2.000000 M 1 Ω 1 µA 0.11 % + 300 0.049 % + 100
20.00000 M 10 Ω 1 µA 0.11 % + 1 k 0.052 % + 500
200.0000 M 100 100 nA 0.655 % + 10 k 0.349 % + 5 k
> 200.0000 M
8
------ User-defined Source I
ACC
+ Meas V
ACC
Meas I
ACC
+ Meas V
ACC
Temperature coefficient: ± (0.15 × accuracy specification)/°C
0 °C to 18 °C and 28 °C to 50 °C
Source current, measure
resistance mode
Total uncertainty = I source accuracy + V measure accuracy (4-wire remote sense)
Source voltage, measure
resistance mode
Total uncertainty = V source accuracy + I measure accuracy (4-wire remote sense)
Guard output impedance
0.5 (DC) in ohms mode
SUPPLEMENTAL SPECIFICATIONS
Overrange
105 % of range, source, and measure
Regulation Voltage
Line: 0.01 % of range
Load: 0.01 % of range + 100 µV
Current
Line: 0.01 % of range
Load: 0.01 % of range + 100 pA
Source limits Voltage source current limit:
Bipolar current limit set with a single value
Minimum value is 10 % of range
Current source voltage limit:
Bipolar voltage limit set with a single value
Minimum value is 10 % of range
V-limit/I-limit accuracy
Add 0.3 % of setting and ±0.02 % of reading to base specification
6
6.5-digit measure resolution.
7
Source readback enabled; offset compensation on.
8
Source current, measure resistance or source voltage, measure resistance only.