Datasheet
Table Of Contents

Model 2450 Interactive SourceMeter
®
Instrument Specifications
Specifications are subject to change without notice
6 of 7 SPEC-2450 Rev. A / July 2013
GENERAL CHARACTERISTICS
(Default mode unless specified)
Factory default standard
power-up setting
SCPI mode
Source output modes
Fixed DC level
Memory/configuration list (mixed function)
Stair (linear and logarithmic)
Source memory list
100 points maximum (SCPI 2400 command set only)
Memory buffer
> 250,000 readings with selected measured values and timestamp
Real-time clock
Lithium battery backup (3 years plus battery life)
Remote interfaces GPIB: IEEE Std 488.1 compliant; supports IEEE Std 488.2 common commands and
status model topology
USB device (rear panel, type B): 2.0 full-speed USBTMC
USB host (front panel, type A): USB 2.0, support for flash drives, FAT32
Ethernet: RJ-45 connector, 10/100 BT
IP configuration
Static or DHCP
Expansion interface
The TSP-Link
®
expansion interface allows TSP-enabled instruments to trigger and
communicate with each other
LXI compliance
LXI version 1.4 Core 2011
TSP mode
Embedded Test Script Processor (TSP) accessible from any host interface
Display
Five-inch capacitive touch, color TFT WVGA (800 × 480) with LED backlight
Input signal connections Front: Banana
Rear: Triaxial (3-lug)
Programmability
SCPI or TSP command sets
Interlock
Active high-input
Digital I/O Lines
Six input/output, user-defined, for digital
I/O or triggering
Connector
9-pin female D
Input signal levels
0.7 V (maximum logic low), 3.7 V
(minimum logic high)
Input voltage limits
-0.25 V (absolute minimum), +5.25 V
(absolute maximum)
Maximum source current
+2.0 mA at > 2.7 V (per pin)
Maximum sink current
-50 mA at 0.7 V (per pin, solid-state fuse
protected)
5 V power supply pin
Limited to 500 mA at > 4 V (solid-state
fuse protected)
Handler
User-definable start of test, end of test,
four category bits