Datasheet

SMU INSTRUMENTS
A Greater Measure of Confidence
www.keithley.com
1.888.KEITHLEY
(U.S. only)
2460
SourceMeter
®
SMU Instrument
100 Watt, 7 Amps
Voltage Specifications
1, 2
Source Measure
3
Range
Max.
Current
Reso-
lution
Accuracy
(23° ±C), 1 Year
±(% setting+volts)
Noise
(RMS)
(<10Hz)
Reso-
lution
4
Input
Resistance
Accuracy
(23° ±C), 1 Year
±(% rdg. + volts)
200.0000 mV 7.35 A 5 μV 0.015 % + 200 μV 1 μV 100 nV >10 GW 0.012 % + 200 μV
2.000000 V 7.35 A 50 μV 0.015 % + 300 μV 10 μV 1 μV >10 GW 0.012 % + 300 μV
7.000000 V 7.35 A 250 μV 0.015 % + 2.4 mV 100 μV 1 μV >10 GW 0.015 % + 1 mV
10.00000 V 5.25 A 500 μV 0.015 % + 2.4 mV 100 μV 10 μV >10 GW 0.015 % + 1 mV
20.00000 V 4.20 A 500 μV 0.015 % + 2.4 mV 100 μV 10 μV >10 GW 0.015 % + 1 mV
100.0000 V 1.05 A 2.5 mV 0.015 % + 15 mV 1 mV 100 μV >10 GW 0.015 % + 5 mV
Current Specifications
1, 2, 5
Source Measure
3
Range
Max.
Voltage
Reso-
lution
Accuracy
(23° ±C), 1 Year
±(% setting + amps)
Noise
(RMS)
(<10Hz)
Reso-
lution
4
Voltage
Burden
6
Accuracy
(23° ±C), 1 Year
±(% rdg. + amps)
1.000000 μA 105 V 50 pA 0.025 % + 1 nA 40 pA 10 pA <100 μV 0.025 % + 700 pA
10.00000 μA 105 V 500 pA 0.025 % + 1.5 nA 40 pA 10 pA <100 μV 0.025 % + 1 nA
100.0000 μA 105 V 5 nA 0.020 % + 15 nA 100 pA 100 pA <100 μV 0.020 % + 10 nA
1.000000 mA 105 V 50 nA 0.020 % + 150 nA 1 nA 1 nA <100 μV 0.020 % + 100 nA
10.00000 mA 105 V 500 nA 0.020 % + 1.5 μA 10 nA 10 nA <100 μV 0.020 % + 1 μA
100.0000 mA 105 V 5 μA 0.020 % + 15 μA 100 nA 100 nA <100 μV 0.020 % + 10 μA
1.000000 A 105 V 50 μA 0.050 % + 750 μA 5 μA 1 μA <100 μV 0.050 % + 500 μA
4.000000 A 21 V 250 μA 0.100 % + 3 mA 25 μA 1 μA <100 μV 0.100 % + 2.5 mA
5.000000 A 10.5 V 250 μA 0.100 % + 3 mA 25 μA 1 μA <100 μV 0.100 % + 2.5 mA
7.000000 A 7.35 V 500 μA 0.150 % + 6 mA 125 μA 1 μA <100 μV 0.150 % + 5 mA
TEMPERATURE COEFFICIENT (0°–18°C and 28°–50°C): ±(0.10 × accuracy specification)/°C.
1. Speed = 1 PLC.
2. All specifications are guaranteed with output ON.
3. Accuracies apply to 2- and 4-wire mode when properly zeroed.
4. 6.5-digit measure resolution.
5. Accuracy specifications guaranteed when using 2460-KIT screw terminal accessory.
6. Four-wire mode.
Resistance Measurement Accuracy (Local or Remote Sense)
7, 8, 9
Range
Default
Resolution
10
Default
Test Current
Normal Accuracy
(23°C ±5°C)
1 Year, ±(% rdg. + ohms)
Enhanced Accuracy
11
(23°C ±5°C)
1 Year, ±(% rdg. + ohms)
<2.000000 W
12
1 μW User defined Source I
ACC
+ Meas. V
ACC
Meas. I
ACC
+ Meas. V
ACC
20.00000 W 10 μW 100 mA 0.05 % + 0.003 W 0.04 % + 0.001 W
200.0000 W 100 μW 10 mA 0.05 % + 0.03 W 0.04 % + 0.01 W
2.000000 kW 1 mW 1 mA 0.05 % + 0.3 W 0.04 % + 0.1 W
20.00000 kW 10 mW 100 μA 0.05 % + 3 W 0.04 % + 1 W
200.0000 kW 100 mW 10 μA 0.05 % + 30 W 0.05 % + 10 W
2.000000 MW 1 W 10 μA 0.06 % + 100 W 0.06 % + 50 W
20.00000 MW 10 W 1 μA 0.14 % + 1000 W 0.12 % + 500 W
>20.0000 MW
12
User defined Source I
ACC
+ Meas. V
ACC
Meas. I
ACC
+ Meas. V
ACC
TEMPERATURE COEFFICIENT (0°–18°C and 28°–50°C): ±(0.10 × accuracy specification)/°C.
SOURCE CURRENT, MEASURE RESISTANCE MODE:
Total uncertainty = I source accuracy + V measure accuracy (4-wire remote sense).
SOURCE VOLTAGE, MEASURE RESISTANCE MODE:
Total uncertainty = V source accuracy + I measure accuracy (4-wire remote sense).
7. Speed = 1 PLC.
8. All specifications are guaranteed with output ON.
9. Accuracies apply to 2- and 4-wire mode when properly zeroed.
10. 6.5-digit measure resolution.
11. Source readback enabled. Offset compensation ON.
12. Source current, measure resistance or source voltage, measure resistance only.
ACCESSORIES AVAILABLE
TEST LEADS AND PROBES
1754 2-wire Universal 10-Piece Test Lead Kit
5805 Kelvin (4-Wire) Spring-Loaded Probes
5808 Low Cost Single-pin Kelvin Probe Set
5809 Low Cost Kelvin Clip Lead Set
8605 High Performance Modular Test Leads
8606 High Performance Modular Probe Kit
8608 High Performance Clip Lead Set
CABLES, CONNECTORS, ADAPTERS
2460-BAN Screw Terminal Connector to Banana Cable
2460-KIT Mating Mass Termination Connector
8607 2-Wire, 1000V Banana Cables, 1m (3.3 ft.)
CS-1616-3 Safety Interlock Mating Connector
COMMUNICATION INTERFACES & CABLES
7007-1 Shielded GPIB Cable, 1m (3.3 ft)
7007-2 Shielded GPIB Cable, 1m (6.6 ft)
CA-180-3A CAT5 Crossover Cable for TSP-Link/Ethernet
KPCI-488LPA IEEE-488 Interface for PCI Bus
KUSB-488B IEEE-488 USB-to-GPIB Interface Adapter
USB-B-1 USB Cable, Type A to Type B, 1m (3.3 ft)
TRIGGERING AND CONTROL
2450-TLINK DB-9 to Trigger Link Connector Adapter.
8501-1 Trigger Link Cable, DIN-to-DIN, 1m (3.3 ft)
8501-2 Trigger Link Cable, DIN-to-DIN, 2m (6.6 ft)
RACK MOUNT KITS
4299-8 Single Fixed Rack Mount Kit
4299-9 Dual Fixed Rack Mount Kit
4299-10 Dual Fixed Rack Mount Kit. Mount one 2460
and one Series 26xxB
4299-11 Dual Fixed Rack Mount Kit. Mount one 2460
and one Series 2400, Series 2000, etc.
2450-BenchKit Ears and Handle for 2460-NFP-RACK and
2460-RACK models
SERVICES AVAILABLE
2460-3Y-EW 1 Year Factory Warranty extended to 3 years
from date of shipment
2460-5Y-EW 1 Year Factory Warranty extended to 5 years
from date of shipment
C/2460-3Y-17025 KeithleyCare
®
3 Year ISO 17025 Calibration Plan
C/2460-3Y-DATA KeithleyCare 3 Year Calibration w/Data Plan
C/2460-3Y-STD KeithleyCare 3 Year Std. Calibration Plan
C/2460-5Y-17025 KeithleyCare 5 Year ISO 17025 Calibration Plan
C/2460-5Y-DATA KeithleyCare 5 Year Calibration w/Data Plan
C/2460-5Y-STD KeithleyCare 5 Year Std. Calibration Plan
C/NEW DATA Calibration Data for New Units
C/NEW DATA ISO ISO-17025 Calibration Data for New Units
Model 2460 specifications
Model 2460 specifications