Datasheet

SMU INSTRUMENTS
A Greater Measure of Confidence
www.keithley.com
1.888.KEITHLEY
(U.S. only)
2460
SourceMeter
®
SMU Instrument
100 Watts, 7 Amps
GENERAL CHARACTERISTICS (default mode unless specified)
FACTORY DEFAULT STANDARD POWER-UP: SCPI MODE.
SOURCE OUTPUT MODES: Fixed DC Level, Memory/Configuration List (mixed function),
Sweep (linear and logarithmic), Sweep (dual linear and dual logarithmic.
MEMORY BUFFER: >250,000 readings. Includes selected measured value(s) and time stamp.
REAL-TIME CLOCK: Lithium battery backup (3 yr. + battery life).
REMOTE INTERFACES:
GPIB: IEEE-488.1 compliant. Supports IEEE-488.2 common commands and status
model topology.
USB Device (rear panel, type B): 2.0 Full Speed USBTMC.
USB Host (front panel, type A): USB 2.0, support for flash drives, FAT32.
Ethernet: RJ-45 (10/100BT)
DIGITAL I/O INTERFACE:
Lines: 6 Input/Output user defined for digital I/O or triggering
Connector: 9-pin female D
Input Signal Levels: 0.7V (maximum logic low), 3.7V (minimum logic high)
Input Voltage Limits: 0.25V (Abs. minimum), +5.25V (Abs. maximum)
Maximum Source Current: +2.0mA @ >2.7V (per pin)
Maximum Sink Current: 50mA @ 0.7V (per pin, solid-state fuse protected)
5V Power Supply Pin: Limited to 500mA @ >4V (solid-state fuse protected)
Handler: User definable Start of Test, End of Test, 4 category bits
PROGRAMMABILITY: SCPI or TSP command sets.
TSP MODE: Embedded Test Script Processor (TSP) accessible from any host interface.
IP CONFIGURATION: Static or DHCP
EXPANSION INTERFACE: The TSP-Link expansion interface allows TSP enabled instruments
to trigger and communicate with each other.
LXI COMPLIANCE: 1.4 LXI Core 2011.
DISPLAY: 5 inch capacitive touch, color TFT WVGA (800×480) with LED backlight.
INPUT SIGNAL CONNECTIONS: Front: Banana. Rear: Mass termination screw terminal.
INTERLOCK: Active High Input.
COOLING: Forced air, variable speed.
OVER TEMPERATURE PROTECTION: Internally sensed temperature overload puts unit in
standby mode.
POWER SUPPLY: 100V to 240V RMS, 5060Hz (automatically detected at power up).
VA RATING: 350 volt-amps max.
ALTITUDE: Maximum 2000 meters above sea level.
EMC: Conforms to European Union EMC Directive.
SAFETY: Compliance with CE and NRTL listed to UL61010-1 and UL61010-2-30. Conforms with
European Union Low Voltage Directive.
VIBRATION: MIL-PRF-28800F Class 3 Random.
WARM-UP: 1 hour to rated accuracies.
DIMENSIONS: (With handle and bumpers): 106mm high × 255mm wide × 425mm deep
(4.18 in × 10.05 in × 16.75 in). (Without handle and bumpers): 88mm high × 213mm wide
× 397mm deep (3.46 in × 8.39 in × 15.63 in.)
WEIGHT: With bumpers and handle: 4.75 kg (10.5 lbs.).
Without bumpers and handle: 4.35 kg (9.6 lbs.).
ENVIRONMENT: Operating: 0°–50°C, 70% R.H. up to 35°C. Derate 3% R.H./°C, 35°–50°C,
non-condensing. Storage: –25°C to 65°C.
ACCESSORIES SUPPLIED: Test Leads, Mating Mass Terminated Screw Connector, USB Cable,
Ethernet/TSP Cable, Interlock Adapter, Power Cord, QuickStart Guide, CD Users Manual.
Source and sink (4-quadrant) operation
Model 2460 specifications
Model 2460 specifications