Datasheet
Datasheet
TEK.COM4
Specification Conditions
This document contains specifications and supplemental information for the 2606B System SourceMeter
®
instrument. Specifications are
the standards against which the 2606B instruments are tested. Upon leaving the factory, the 2606B instruments meet these specifications.
Supplemental and typical values are nonwarranted, apply at 23°C, and are provided solely as useful information. Specifications are for
individual modules.
Source and measurement accuracies are specified at the 2606B terminals under these conditions:
1. 23°C ±5°C, <70 percent relative humidity
2. After a two-hour warm-up period
3. Speed normal (1 NPLC)
4. A/D autozero enabled
5. Remote sense operation or properly zeroed local operation
6. Calibration period: One year
DC Power Specifications
Voltage Current
Maximum output power
and source limits
1, 2
20.2 W maximum
20.2 V at 1.0 A, –20.2 V at –1.0 A
6.06 V at 3.0 A, –6.06 V at –3.0 A
20.2 W maximum
1.01 A at 20 V, –1.01 A at –20 V
3.03 A at 6 V, –3.03 A at –6 V
+3 A
+1 A
–3 A
–1 A
+20 V
DC & Pulse
–20 V 0 V
0 A
+6 V–6 V
Voltage Accuracy Specifications
3, 4
Range
Source Measure
Programming
Resolution
Accuracy
±(% reading + volts)
Typical Noise (peak to peak)
0.1 Hz to 10 Hz
Display
Resolution
Accuracy
±(% reading + volts)
100 mV 5 µV 0.02% + 250 µV 20 µV 100 nV 0.015% + 150 µV
1 V 50 µV 0.02% + 400 µV 50 µV 1 µV 0.015% + 200 µV
6 V 50 µV 0.02% + 1.8 mV 100 µV 1 µV 0.015% + 1 mV
20 V 500 µV 0.02% + 12 mV 500 µV 10 µV 0.015% + 8 mV
Notes
1. For additional power derating information for various load and temperature conditions, refer to “Operating Boundaries” in the Series 2606B Reference Manual.
2. The System SourceMeter
®
will allow quadrant II and quadrant IV operation, however, it is intended for low power or short transient behavior. Operation in quadrant II and quadrant IV could
result in an overtemperature error.
3. For temperatures 0°C to 18°C and 28°C to 50°C, accuracy is degraded by ±(0.15 × accuracy specification)/°C.
4. Add 50 µV to source accuracy specifications per volt of HI lead drop.