Datasheet

Datasheet
TEK.COM6
Pulse Width Programming Resolution
1 µs
Pulse Width Programming Accuracy
±5 µs
Pulse Width Jitter 2 µs
Transient Response Time <70 µs for the output to recover to within 0.1% for a 10% to 90% step change in load.
Overshoot Voltage: <±0.1% of range + 10 mV. Step size = 10% to 90% of range, resistive load, maximum current
limit/compliance
Current:0.1% of range. Step size = 10% to 90% of range, resistive load. See Current Source Output
Settling Time for additional test conditions.
Range Change Overshoot Voltage: <300 mV + 0.1% of larger range. Overshoot into a 100 kΩ load, 20 MHz bandwidth.
Current
7
: <300 mV/R
LOAD
+ 5% of larger range
Guard Offset Voltage <4 mV, current <10 mA
Remote Sense Operating Range
8
Maximum voltage between HI and SENSE HI = 3 V
Maximum voltage between LO and SENSE LO = 3 V
Voltage Output Headroom 20 V Range: Maximum output voltage = 22 V – (total voltage drop across source leads). Maximum 1 Ω
source lead.
6 V Range: Maximum output voltage = 8 V – (total voltage drop across source leads). Maximum 1 Ω
source lead.
Overtemperature Protection Internally sensed temperature overload puts the instrument in standby mode
Limit/Compliance Bipolar limit (compliance) set with a single value.
Voltage: Minimum value is 10 mV; accuracy is the same as voltage source.
Current: Minimum value is 10 nA; accuracy is the same as current source
Voltage Source Output Settling Time
Time required to reach within 0.1% of final value after source level command is processed on a fixed range.
Voltage Range Settling Time
100 mV < 50 µs
1 V < 50 µs
10 V < 110 µs
20 V
9
< 150 µs
Current Source Output Settling Time
Time required to reach within 0.1% of final value after source level command is processed on a fixed range.
Values below for I
OUT
× R
LOAD
= 1 V
Current Range Settling Time
3 A
< 80 µs (Current < 2.5 A, R
LOAD
> 2 Ω)
1 A to 10 mA
< 80 µs (R
LOAD
> 6 Ω)
1 mA < 100 µs
100 µA < 150 µs
10 µA < 500 µs
1 µA < 2 ms
100 nA < 20 ms
Notes
7. With source settling set to SETTLE_SMOOTH_100NA
8. Add 50 µV to source accuracy specifications per volt of HI lead drop.
9. Add 150 µs when measuring on the 1 A range.