Datasheet
Datasheet
TEK.COM8
Maximum Single Measurement Rates (operations per second) for 60 Hz (50 Hz)
A/D Converter Speed
(NPLC) Trigger Origin Measure to USB Source Measure to USB
Source Measure
Pass/Fail to USB
0.001 Internal 1900 (1800) 1400 (1400) 1400 (1400)
0.01 Internal 1450 (1400) 1200 (1200) 110 0 (110 0)
0.1 Internal 450 (390) 425 (370) 425 (375)
1.0 Internal 58 (48) 57 (48) 57 (48)
Maximum Measurement Range Change Rate
>7000 per second for >10 µA. When changing to or from a range ≥1 A, maximum rate is >2200/second.
Maximum Source Range Change Rate
>400 per second >10 µA. When changing to or from a range ≥1 A, maximum rate is >190/second.
Maximum Source Function Change Rate
>1000 per second
Command Processing Time <1 ms. Maximum time required for the output to begin to change after receiving the
smua.source.levelv or smua.source.leveli command.
Triggering and Synchronization Characteristics
Triggering
Trigger In to Trigger Out 0.5 µs
Trigger In to Source Change
13
10 µs
Trigger Timer Accuracy ±2 µs
Source Change After LXI Trigger 280 µs
Synchronization
Multi-node Synchronized Source Change
13
< 0.5 µs
Single-node Synchronized Source Change
13
< 0.5 µs
Notes
13. Fixed source range with no polarity change