This setting controls the aperture of measurements made for contact check. It does not affect the 
smuX.measure.nplc aperture setting. 
The speed setting can have a dramatic effect on the accuracy of the measurement (see specifications). 
Example 
smua.contact.speed = smua.CONTACT_SLOW 
Configure contact check for higher accuracy 
on SMU channel A. 
Also see 
Contact check connections (on page 2-54) 
Contact check measurements (on page 2-44) 
reset() (on page 7-160) 
smuX.contact.check() (on page 7-193) 
smuX.contact.r() (on page 7-194) 
smuX.reset() (on page 7-216) 
smuX.contact.threshold 
This attribute stores the resistance threshold for the smuX.contact.check() function. This attribute is not 
available on the Models 2604B/2614B/2634B. 
Type  TSP-Link accessible  Affected by  Where saved  Default value 
Attribute (RW) 
Yes 
Instrument reset 
SMU reset 
Recall setup 
Not saved 
50 (50 Ω) 
Usage 
rValue = smuX.contact.threshold 
smuX.contact.threshold = rValue 
rValue 
The resistance above which contact check should fail (measured in ohms) 
X 
Source-measure unit (SMU) channel (for example, smua.contact.threshold 
applies to SMU channel A) 
Details 
The threshold should be set to less than 1 kΩ. 
Example 
smua.contact.threshold = 5 
Set the contact check threshold for SMU channel A to 5 Ω. 
Also see 
Contact check connections (on page 2-54) 
Contact check measurements (on page 2-44) 
reset() (on page 7-160) 
smuX.contact.check() (on page 7-193) 
smuX.reset() (on page 7-216) 