Operation Manual
7-6 Scanning Model 2750 Multimeter/Switch System User’s Manual
Figure 7-2
Trigger model with SCAN function
Enable Scan
Control
Source
Immediate
External
Timer
Manual*
Bus*
Event
Detection
Another
Scan?
Trigger
Counter
Yes
No
Close First
Chan in List
Open Last Chan
Close Next Chan
in List
Ratio/Chan
Average Delay
Measurement
Process
Device Action
Output
Trigger
Another
Reading
?
Yes
No
Sample
Counter
*Remote programming only
Yes
Timer >
Delay
?
Timer
Enabled
?
Yes
No
Timer
Bypass
No
Delay
(Auto or Manual)
Timer
2750-900-01.book Page 6 Wednesday, August 3, 2011 7:56 AM