User manual

Series 3700 System Switch/Multimeter Reference Manual Section 14: Verification
3700S-901-01 Rev. C / July 2008 14-23
Figure 14-11: Resistance verification
Dry circuit resistance verification data
Use the following values to verify the performance of the Series 3700. Actual
values depend on published specifications (see Calculating resistance reading
limits (on page 14-4)).
Description
Range (Ohms)
Test point
(Ohms)
Lower limit (Ohms)
Upper limit
(Ohms)
Verify Dry Circuit 100
Ohm *
1.00E+02
1.00E+02
9.997800E+01
1.000220E+02
Verify Dry Circuit 1k Ohm
1.00E+03
1.00E+03
9.995200E+02
1.000480E+03
Verify Dry Circuit 2k Ohm
2.00E+03
1.90E+03
1.898320E+03
1.901680E+03
NOTE The asterisk (*) designates the ranges that offset compensation is
being used.