2-4 Connections Model 6485 and 6487 User’s Manual
Voltage source output connectors
The rear panel V-SOURCE OUTPUT HI and LO connectors (see Figure 1-4 in Section 1)
are used to connect the voltage source to the DUT. The voltage source is primarily used
for ohms measurements. See “Ohms measurement connections,” page 2-7 for details.
Maximum input levels
The maximum input levels to the Model 6487 are summarized in Figure 2-4.
t^okfkd The maximum safe voltage between the voltage source or ammeter com-
mon and chassis ground (common mode voltage) is 505V peak. Exceeding
this voltage can create a shock hazard.
Figure 2-4
Model 6487 maximum input levels
Connecting cables and test leads
When making precision measurements, you should always use low-noise triax cables for
INPUT connections. As a general rule, always use the shortest possible cable for measure-
ments. When using the voltage source, the test leads must be rated for 505V minimum and
should include safety sheaths. See “Model 6487 input cables, connectors, and adapters,”
page 1-5 in for recommended cables and test leads.
Input HI
Input LO
505V Peak
Chassis Ground
Max Continuous
Input = 505V Peak
505V Peak