Datasheet

Model DMM7510
7-1/2 Digit Graphical Sampling Multimeter
Specifications
Specifications are subject to change without notice.
SPEC-DMM7510 Rev. B / October 2016 Page 1 of 25
Keithley Instruments
28775 Aurora Road
Cleveland, Ohio 44139
1-800-935-5595
www.tek.com/keithley
SPECIFICATION CONDITIONS
This document contains specifications and supplemental information for the Model DMM7510 7-1/2 Digit
Graphical Sampling Multimeter instrument. Specifications are the standards against which the DMM7510 is
tested. Upon leaving the factory, the DMM7510 meets these specifications. Supplemental and typical
values are nonwarranted, apply at 23 °C (73 °F), and are provided solely as useful information.
Measurement accuracies are specified at the DMM7510 terminals under these conditions:
Temperature 23 °C ±5 °C, 5 % to 80 % relative humidity, noncondensing
After a 90-minute warmup period
1 PLC or 5 PLC; for NPLC settings less than 1 PLC, add appropriate ppm of range for peak noise
uncertainty from the RMS noise table
Autozero enabled unless otherwise noted
Remote sense operation or properly zeroed local operation
Calibration period: One year or two years (calibration period may vary depending on customer
requirements)
T
ACAL
= Ambient temperature of last automatic calibration
T
CAL
= Ambient temperature of last external calibration; factory calibration performed at 23 °C ±1 °C
DC VOLTAGE
ACCURACY (INPUT IMPEDANCE AUTO)
Range
1
Resolution Input
impedance
Accuracy ±[ppm of reading + ppm of range]
24 hour
T
CAL
±1 °C
2
90 day
T
CAL
±5 °C
1 year
T
CAL
±5 °C
2 year
T
CAL
±5 °C
Temperature
coefficient
3
100.00000 mV
4
10 nV
> 10 G or
10 M ±1 %
6 + 9 12 + 9 18 + 9 29 + 9 0.1 + 2.5
1.0000000 V
4
100 nV
> 10 G or
10 M ±1 %
4 + 1 9 + 2 15+ 2 26 + 2 0.1 + 0.5
10.000000 V
4
1 µV
> 10 G or
10 M ±1 %
2 + 0.7 9 + 1.2 14 + 1.2 22 + 1.2 0.1 + 0.05
100.00000 V
4
10 µV
10 M ±1 %
8 + 3 [18 + 5]
5
[22 + 5]
5
[30 + 5]
5
[0.15 + 0.05]
5
35 + 5 40 + 5 45 + 5 2.0 + 0.5
1000.0000 V
4,6
100 µV
10 M ±1 %
8 + 3 [19 + 5]
5
[23 + 5]
5
[31 + 5]
5
[0.15 + 0.05]
5
35 + 5 40 + 5 45 + 4 2.0 + 0.5
1
20 % overrange on all ranges except 1 % for 1000 V range.
2
Relative to calibration accuracy.
3
Add per degree from T
CAL
± 5
o
C.
4
When properly zeroed using the Rel function with external cables.
5
Specified within 30 days of autocalibration, T
OPER
± 5
o
C from T
ACAL
.
6
For signal levels greater than 500 V, add 0.02 ppm/V to the ppm of the readings specification for measurements exceeding 500 V.