6-8-2 While holding the mounting lever of the
filter holder, insert a filter.
(1) Hold the filter by its edge to avoid leaving
fingerprints or smudges on the filter surfaces.
(2) After the illumination has been ignited, the
filter will be very hot. Be sure to set the main
switch to "O"(OFF) and allow the filter holder and
filters to cool down before replacing filters.
6-8-3 Engage each filter in the light path by moving
the filter holder in the direction of the arrow.
6-9 Attaching the Photomicrographic System
6-9-1 This is an optional part for the
photomicrograph.
6-9-2 Using the Allen screwdriver, loosen the
clamping screw and remove the cap.
6-9-3 Align the index of the straight photo tube with
the positioning index on the side port and fit the
straight photo tube.
6-9-4 Tighten the clamping screw firmly.
6-9-5 When the side port is not used, attach the cap for
protecting it from dust.
6-10 Attaching the TV Observation System
6-10-1 This is an optional part for TV observation.
6-10-2 Attaching the photographic eyepiece and
adaptor on the side port.
6-10-3 Connect the CCD camera with screw thread
to get images that art brighter and higher in
resolution.
6-10-4 For the TV adaptor systems, refer to the instruction
manual for the TV adaptor to be used.
7. Preparations for the use of special devices
7-1 Attaching the fluorescence device of inverted
microscope. (Please refer to the specifications for
installation of Fluorescence device.)
7-2 Attaching the polarization device of inverted
microscope. (Please refer to the specifications for
installation of polarization device.)
7-3 Attaching the patch clamp, thermostatic
cover and the thermostatic stage.
7-3-1 Remove the cover of the 4 positions, and link
the patch clamp, thermostatic cover, and the thermostatic
stage to the microscope with the special screws provided by our company.