KERN & Sohn GmbH Ziegelei 1 D-72336 Balingen email: info@kern-sohn.com Phone: +49-[0]7433- 9933-0 Fax: +49-[0]7433-9933-149 Internet: www.kern-sohn.com Operating instructions Analytical balance KERN ABS-N_ABJ-NM Version 1.
GB KERN ABS-N_ABJ-NM Version 1.2 03/2013 Operating instructions Analytical balance Table of Contents 1 Technical Specifications ............................................................................... 4 2 Declaration of conformity ............................................................................. 7 3 Appliance overview ....................................................................................... 8 3.1 3.1.1 3.2 Keyboard overview ..........................................
10.7.1 Freely programmable weighing unit ..................................................................................... 37 10.7.2 Change position of decimal dot for freely programmable weighing unit ............................... 38 10.7.3 Input minimum weight for freely programmable weighing unit ............................................. 38 10.8 Change readability (1D/10D) .................................................................................................. 39 10.
1 Technical Specifications KERN ABJ 80-4NM ABJ 120-4NM ABJ 220-4NM ABJ 320-4NM Readability (d) 0.1 mg 0.1 mg 0.1 mg 0.1 mg 80 g 120 g 220 g 320 g Minimum load (Min) 10 mg 10 mg 10 mg 10 mg Verification value (e) 1 mg 1 mg 1 mg 1 mg I I I I 0.2 mg 0.2 mg 0.2 mg 0.2 mg ± 0.3 mg ± 0.3 mg ± 0.3 mg ± 0.3 mg Weighing range (max) Verification class Reproducibility Linearity Stabilization time 3 sec.
KERN ABS 80-4N ABS 120-4N 0.1 mg 0.1 mg 80 g 120 g 0.2 mg 0.2 mg Linearity ± 0.3 mg ± 0.
KERN ABS 220-4N ABS 320-4N Readability (d) 0.1 mg 0.1 mg Weighing range (max) 220 g 320 g Reproducibility 0.2 mg 0.2 mg ± 0.3 mg ± 0.
2 Declaration of conformity KERN & Sohn GmbH D-72322 Balingen-Frommern Postbox 4052 email: info@kern-sohn.de Phone: 0049-[0]7433-9933-0 Fax: 0049-[0]7433-9933-149 Internet: www.kern-sohn.
3 Appliance overview 1. Weighing plate 2. Display 3. Keyboard 4. Levelling screw 5. Bubble level 6. Name plate 7. Wind protection 8. Appliance interface 9.
3.1 Keyboard overview In menu: Function Key Designation: ON/OFF CAL TARE UNIT Navigation key Pressed once and released Switches between the operation and standby modes. ABS-N_ABJ-NM-BA-e-1312 - Menu item back. Return to weighing mode: Press ON/OFF repeatedly or for 3 s. Cancel procedure Invoke calibration menu Tare or set weight display to zero Invoke zero / tare menu Confirm entry Switch-over weighing unit. Display stored reference weight (PCS, %). Invoke unit setting menu.
3.1.1 Numeric entry Key Designation: Navigation key Navigation key Function Increase flashing digit Shift flashing decimal dot to the left Decrease flashing digit Shift flashing decimal dot to the right Navigation key Digit selection to the right Navigation key Confirm entry ESC Cancel input With numerical input the indicator [#] appears.
3.2 Overview of displays Symbols Display Numerical display/menu Units Description see chpt. Battery symbol Zero tracking function active Displays adjustment Flashes prior to start of automatic adjustment, only models ABJ Stability and reaction settings via „Easy Setting display“ Balance is in dosing mode (pouring mode) Balance is in formula mode Menu lock active Menu symbol Auto Print function active + + + + + + + + + Chpt.12.1 Chpt. 8.2.1 Chpt. 8.1 Chpt.13.1 Chpt. 13.2 Chpt. 14.3 Chpt. 11.5 Chpt.
4 Basic Information (General) 4.1 Proper use The balance you purchased is intended to determine the weighing value of material to be weighed. It is intended to be used as a “non-automatic“ balance, i.e. the material to be weighed is manually and carefully placed in the centre of the weighing plate. As soon as a stable weighing value is reached the weighing value can be read. 4.
4.4 Monitoring of Test Resources In the framework of quality assurance the measuring-related properties of the balance and, if applicable, the testing weight, must be checked regularly. The responsible user must define a suitable interval as well as type and scope of this test. Information is available on KERN’s home page (www.kern-sohn.com with regard to the monitoring of balance test substances and the test weights required for this.
6.2 Packaging / return transport Keep all parts of the original packaging for a possibly required return. Only use original packaging for returning. Prior to dispatch disconnect all cables and remove loose/mobile parts. Reattach possibly supplied transport securing devices. Secure all parts such as the glass wind screen, the weighing platform, power unit etc. against shifting and damage.
ABS-N_ABJ-NM-BA-e-1312 15
7 Unpacking, Setup and Commissioning 7.1 Installation Site, Location of Use The balances are designed in a way that reliable weighing results are achieved in common conditions of use. You will work accurately and fast, if you select the right location for your balance. Therefore, observe the following for the installation site: • Operate the device only indoors.
7.2.1 Scope of delivery / serial accessories 1. Balance 2. Weighing plate 3. Carrier weighing plate 4. Screening ring 5. Mains adapter 6. Operating instructions 7.
7.2.2 Placing Attach circular screen, carrier of weighing plate and weighing plate in order. Level balance with foot screws until the air bubble of the water balance is in the prescribed circle.
7.3 Mains connection Power is supplied via the external mains adapter. The stated voltage value must be the same as the local voltage. Only use original KERN mains adapters. Using other makes requires consent by KERN. 7.3.1 Turning On the Power Supply power to balance via mains adapter. The display lights up and the balance carries out a selftest. In the ABJ models an automatic adjustment takes place. The selftest is completed when „OFF“ appears on the display. 7.
8 Adjustment As the acceleration value due to gravity is not the same at every location on earth, each balance must be coordinated - in compliance with the underlying physical weighing principle - to the existing acceleration due to gravity at its place of location (only if the balance has not already been adjusted to the location in the factory). This adjustment process must be carried out for the first commissioning, after each change of location as well as in case of fluctuating environment temperature.
Case 1: The load on the weighing plate is near zero. The weight symbol flashes for approx. two minutes, followed by „PSC.RUN“. After that the internal adjustment is started automatically. In order to ensure proper PSC operation, prevent vibrations and air flow. As soon as the display in grams reappears after completing adjustment via PSC, the balance returns to weighing mode.
Case 2: Weighing plate loaded The gram display flashes for two minutes, followed by „PLS.CAL“. Unload weighing plate. The gram display flashes again for 2 minutes; after that the internal adjustment is started automatically. In order to ensure proper PSC operation, prevent vibrations and air flow. As soon as the display in grams reappears after completing adjustment via PSC, the balance returns to weighing mode.
8.2 Menu settings „I.CAL“ / „E.CAL“ To invoke the adjustment function in weighing mode, press CAL for 3 sec. Acknowledge using PRINT, the current setting is displayed. Select the desired adjustment using the navigation keys ( ) I.CAL: E.CAL: Adjustment with internal weight (see chpt. 8.1) Adjustment with external weight (see chpt. 8.2) Confirm with TARE Press ON/OFF repeatedly or 3 sec., the balance will return into weighing mode The saved adjustment (I.CAL or E.
8.2.1 Adjustment with internal weight (KERN ABJ) With the internal adjustment weight, the weighing accuracy can be checked and readjusted at any time. • Condition: Menu setting „I.CAL“. • When an optional printer is connected and the GLP function activated, while starting the adjustment log „WAIT“ will be displayed. After the printout the adjustment will be continued automatically. • Adjustment may be interrupted with ON/OFF, „ABORT“ is displayed.
8.2.2 Adjustment with external weight (KERN ABS) • Condition: Menu setting „E.CAL“. • The weight to be used depends on the capacity of the scale. Carry out adjustment as near as possible to the balance’s maximum weight (recommended adjustment weight see chpt. 1). Weights of different nominal values or tolerance classes may be used for adjustment but are not optimal for technical measuring.
8.3 Adjustment log This function enables automatic log issue after each adjustment. These logs may be issued by using an optional printer. Printout example (KERN YKB-01N): ----------------CAL –EXTERNAL ----------------KERN & Sohn GmbH TYPE ABJ 220-4NM SN WBIIAB000I ID 1234 Mode of adjustment Company Model Serial no. Balance identification no. (see chpt. 8.4) REF 200.0000g Used adjustment weight BFR 200.0001g Before adjustment AFT 200.
How to change settings Press TARE Press ON/OFF repeatedly or for 3 sec., the balance will return into weighing mode 8.4 Balance identification no. This setting is for the balance ID number that is output along with the adjustment report. Call up menu In weighing mode press the MENU button twice Select menu item Press the navigation buttons ( ) repeatedly until „TOOLS“ is displayed. Confirm with PRINT Press the navigation buttons ( ) repeatedly until „SYSTEM“ is displayed.
9 Verification General introduction: According to EU directive 90/384/EEC or 2009/23EG balances must be officially verified if they are used as follows (legally controlled area): a) For commercial transactions if the price of goods is determined by weighing. b) For the production of medicines in pharmacies as well as for analyses in the medical and pharmaceutical laboratory. c) For official purposes d) For manufacturing final packages In cases of doubt, please contact your local trade in standard.
Position of seals Seal Data plate Seal ABS-N_ABJ-NM-BA-e-1312 29
10 Weighing 10.1 Switch-on balance / call-up weighing mode Status balance Display switched off Call-up weighing mode Press ON/OFF. After display press any key Display Display Press any key All segments light up Balance is in menu After numeric input Press ON/OFF repeatedly or for 3 sec. 10.2 Switch off the balance Press ON/OFF. The balance is in standby mode, that means that the balance is now in state ready-foroperation.
10.3 Set the start display For the start display you can choose between three types. Start display Explanation 1. Weighing mode After connection to the power supply the balance will start in weighing mode. 2. Display After connection to the power supply the balance displays . After pressing a certain key the balance will carry out a segment test and start in weighing mode automatically. 3. All segments After connection to the power supply the balance displays .
Call up menu In weighing mode press the MENU button twice Select menu item Press the navigation buttons ( ) repeatedly until „TOOLS“ is displayed. Confirm with PRINT Press the navigation buttons ( ) repeatedly until „SYSTEM“ is displayed. Confirm with PRINT Press the navigation buttons ( ) repeatedly until „START“ is displayed. The current setting is marked by the stability display ( ). Set display type Acknowledge using PRINT, the current setting is displayed.
10.4 Auto Power-Off function (automatic switch-off function) To save battery when function is enabled backlight of display will automatically switch off after a defined time during which there was no change of load or activity. 1. Call up menu In weighing mode press the MENU button twice 2. Select function Press the navigation buttons ( ) repeatedly until „TOOLS“ is displayed. Confirm with PRINT Press the navigation buttons ( ) repeatedly until „SYSTEM“ is displayed.
10.5 Simple weighing A warm-up time is required for stabilisation (see chpt. 1). Wait for zero display, reset to zero using TARE. Place the goods to be weighed and close the wind screen doors Wait until the stability display appears [ ]. Read weighing result. When an optional printer is connected, the weighing value can be edited. Print-out example with enabled GLP function (see chpt. 8.3): ----------------KERN & Sohn GmbH TYPE ABJ 220-4NM SN WBIIAB000I ID 1234 ----------------- 50.
10.6 Taring The dead weight of any weighing container may be tared away by pressing a button, so that the following weighings show the net weight of the goods to be weighed. Put vessel of goods to be weighed on weighing plate and close the wind screen doors. Wait until the stability display appears ( ), then press TARE. The weight of the container is now internally saved. Weigh the goods to be weighed and close the wind screen doors. Wait until the stability display appears ( ). Read net weight.
10.7 Switch-over weighing unit By pressing the UNIT key, the display can be switched over to the units enabled before in the menu. Call up menu In weighing mode press the UNIT key for 3 sec. The current setting is marked by the stability display ( ). With stability display ( ) Unit enabled Without stability display ( ) Unit disabled Enable/disable units Press TARE Use the navigation buttons ( ) to select more units and to enable/disable them as described before.
10.7.1 Freely programmable weighing unit Call up menu In weighing mode press the UNIT key for 3 sec. Use the navigation buttons ( ) to select „UNIT.U“. The current setting is marked by the stability display ( ). With stability display ( ) Unit enabled Without stability display ( ) Unit disabled if necessary enable units Press TARE Enter conversion factor 1. Press TARE and current setting will be displayed. 2. Use the navigation buttons to enter the desired conversion factor (see chapter 3.1.
10.7.2 Change position of decimal dot for freely programmable weighing unit The position of the decimal dot can only be changed by a numeric input of the conversion factor (see chpt. 9.7.1, step 2). When the first cipher flashes, press PRINT repeatedly until the decimal dot flashes. Select the desired position using the navigation keys ( ). If no decimal dot shall be set, press MENU repeatedly until the display symbol appears. Confirm with TARE. Press ON/OFF repeatedly or for 3 sec.
10.8 Change readability (1D/10D) Call up menu In weighing mode press the MENU button twice Select menu item Press the navigation buttons ( ) repeatedly until „TOOLS“ is displayed. Confirm with PRINT Press the navigation buttons ( ) repeatedly until „TARGET“ is displayed. Confirm with PRINT Press the navigation buttons ( ) repeatedly until „PARAM.W“ is displayed. Confirm with PRINT Press the navigation buttons ( ) repeatedly until „CHG.MIN“ is displayed.
10.9 Display decimal dot as point or comma Call up menu In weighing mode press the MENU button twice Select menu item Press the navigation buttons ( ) repeatedly until „TOOLS“ is displayed. Confirm with PRINT Press the navigation buttons ( ) repeatedly until „SYSTEM“ is displayed. Confirm with PRINT Press the navigation buttons ( ) repeatedly until „POINT“ is displayed. Selection point/comma Acknowledge using PRINT, the current setting is displayed.
11 Menu To adapt the behaviour of the balance to your requirements, go to the menu. Usually the default setting of the menu is such that you do not need to make any changes. If you encounter special conditions of use, go to the menu in order to set your balance according to your individual requirements.
11.2 Navigation in the menu • How to invoke a menu, see chpt. 11 • Menu structure • • Select and pass through menu items to bottom (). Select setting within one function • • Select and pass through menu items to top (). Select setting within one function • • After selecting the function in the display via the navigation buttons ( ), the change is invoked by pressing the TARE key. Confirm and store the setting momentarily appearing on the display by actuating the TARE key.
11.
11.4 Resetting the menu This will return all the settings to factory setting. The reference value stored in previous use of piece counting or percentage conversion will also be cleared. Factory settings are marked by a “*” in the menu oversight. Call up menu In weighing mode press the MENU button twice Select menu item Press the navigation buttons ( ) repeatedly until „TOOLS“ is displayed. Confirm with PRINT Press the navigation buttons ( ) repeatedly until „SYSTEM“ is displayed.
11.5 Menu Lock The menu setting operations can be locked so that the settings cannot be inadvertently changed. This menu lock is set with the following procedure. Either Switch-on power supply of the balance and wait until „OFF“ is displayed or or Set balance into stand-by mode, see chpt. 10.2 Select menu item 1. Press MENU for 3 sec., the password inquiry appears. 2. Use the navigation keys ( ) to enter the password see chpt. 3.1.1 „Numeric input“.
11.5.1 Change password Standard password (factory setting): „9999“ Call up menu In weighing mode press the MENU button twice Select menu item Press the navigation buttons ( ) repeatedly until „TOOLS“ is displayed. Confirm with PRINT Press the navigation buttons ( ) repeatedly until „SYSTEM“ is displayed. Confirm with PRINT Press the navigation buttons ( ) repeatedly until „PASS.WRD“ is displayed. Change password Confirm with TARE.
11.6 Log menu settings When an optional printer is connected, a list of the current menu settings can be printed out. + Data output, see chpt. 16 Call up menu In weighing mode press the MENU button twice Select menu item Press the navigation buttons ( ) repeatedly until „TOOLS“ is displayed. Confirm with PRINT Press the navigation buttons ( ) repeatedly until „SYSTEM“ is displayed. Confirm with PRINT Press the navigation buttons ( ) repeatedly until „COND.OUT“ is displayed.
12 Zero / tare menu (zeroing and taring functions) Selectable functions: 1. Zero tracking function + see chpt 12.1 This function is used to correct automatically small weight variations which appear directly after switching-on. In the event that small quantities are removed or added to the material to be weighed, incorrect weighing results can be displayed due to the “stability compensation“ in the balance. (e.g. slow flow of liquids from a container placed on the balance, evaporating processes).
12.1 Zero tracking function The zero tracking function „A.ZERO“ is turned-on by factory. 1. Check menu setting Zero tracking symbol Zero tracking function is displayed switched on is not displayed switched off 2. Call function In weighing mode press TARE for 3 sec. and if necessary press repeatedly the navigation buttons ( ). until „Z.TRC“ is displayed. The current setting is marked by the stability display ( ).
12.2 Auto zero function The auto zero function is not available when formula mode is enabled (see chapter 14.3). 1. Call up menu Press TARE for 3 sec. in weighing mode. 2. Select function Press the navigation buttons ( ) repeatedly until „A.ZERO“ is displayed. The current setting is marked by the stability display ( ). Further steps depend on the desired setting: Stability display ( yes ) Function no ON off set/update cancel Press PRINT and continue with step 3.
12.3 Auto Tare function 1. Call up menu Press TARE for 3 sec. in weighing mode. 2. Select function Press the navigation buttons ( ) repeatedly until „A.TARE“ is displayed. The current setting is marked by the stability display ( ). With stability display ( ) Function turned on Without stability display ( ) Function switched off 3. Enable/disable function Press TARE 4. Return to weighing mode Press ON/OFF repeatedly or for 3 sec.
12.4 Zero / tare timing change function The zero / tare timing change function can be switched-on, when the „Auto zero“ and the „Auto tare“ function are enabled. 1. Call up menu Press TARE for 3 sec. in weighing mode. 2. Select function Press the navigation buttons ( ) repeatedly until „TARE.F“ is displayed. Confirm with PRINT. The current setting is marked by the stability display ( ). With stability display ( ) Function turned on Without stability display ( ) Function switched off 3.
13 Settings for Stability and Response Exists the possibility to tune the stability of the display and the degree of reaction of the balance to the requirements of certain applications or the environmental conditions. Most measurements can be carried out using default settings, that is, standard mode. In standard weighing mode, stability and reaction have the same priority. For certain applications such as e.g. dosage do use the dosing mode (pouring mode).
13.2 Selection weighing / dosing mode Call-up weighing mode: This is the default setting. Use this mode, when neither stability nor reaction time must be increased or reduced. Call up menu In weighing mode press the MENU button twice Select menu item Press the navigation buttons ( ) repeatedly until „STAND“ is displayed. Confirm with TARE. The balance returns automatically into weighing mode. Call-up dosing mode: Use this function if you wish to increase display speed, e.g. during apportioning.
13.3 Standstill width If the stability display lights up ( ), the weighing result will be stable within the range indicated by the standstill width. Set range for stability determination: Call up menu In weighing mode press the MENU button twice Select menu item Press the navigation buttons ( ) repeatedly until „TOOLS“ is displayed. Confirm with PRINT Press the navigation buttons ( ) repeatedly until „PARAMW“ is displayed.
Set reaction time: Call up menu In weighing mode press the MENU button twice Select menu item Press the navigation buttons ( ) repeatedly until „TOOLS“ is displayed. Confirm with PRINT Press the navigation buttons ( ) repeatedly until „PARAMW“ is displayed. Confirm with PRINT Press the navigation buttons ( ) repeatedly until „STB.Mk“ is displayed. Set reaction time Confirm with PRINT. The current setting is marked by the stability display ( ).
14 Application Functions • The application functions can be combined with the check weighing or target function (see chpt. 14). • The balance starts in the mode, in which it has been switched off. • For switching over between application and weighing mode, press MENU for 3 sec. 14.1 Parts counting With parts counting you can either count parts into a container or remove parts from a container.
Set reference Place an empty container on the balance and tare it by pressing TARE. Press MENU twice. Press the navigation buttons ( ) repeatedly until the desired reference quantity is displayed (possible reference quantities 5, 10, 20, 50, 100). Use MENU to scroll forward. Use UNIT to scroll backward. Fill number of pieces into the container according to the selected reference quantity. Wait for the stability display ( ), then confirm by pressing TARE.
Or count more pieces or The menu item to change the stored reference is called up, the currently set reference quantity is displayed. + see „4. Change or add piece weight“ The stored piece weight is displayed in grams, marked by the symbol . Using PRINT the reference weight (UW= Unit weight) can be edited when a printer is connected. e.g.: UW = 1.0001 To go back to piece counting display press UNIT anew. Switching over between piece counting and weighing mode 4.
14.2 Percent determination Percent weighing allows to display weight in percent, in relation to a reference weight. The balance offers two possibilities: 1. Reference = 100 % 2. Reference = defined by user Call up menu In weighing mode press the MENU button twice Select percent weighing mode Press the navigation buttons ( ) repeatedly until „APL.FUNC“ is displayed. Confirm with PRINT Press the navigation buttons ( ) repeatedly until „PERCENT“ is displayed.
14.2.1 Reference weight = 100 % Select percent weighing mode, see chpt. 14.2 Press TARE. If necessary, press the navigation buttons ( ) repeatedly until „SAMPLE“ is displayed. Press TARE. Display, if no reference weight is stored Display, if a reference weight is stored Set reference If necessary, place an empty container on the balance and tare it by pressing TARE. Press MENU twice. Put a reference weight which corresponds to 100 %.
14.2.2 User-defined reference Select percent weighing mode, see chpt. 13.2 Press TARE. If necessary, press the navigation buttons ( ) repeatedly until „OPTION“ is displayed. Press TARE. Display, if no reference is stored Display, if a reference is stored Set reference If necessary, place an empty container on the balance and tare it by pressing TARE. Press MENU twice. Press PRINT and current setting will be displayed. Enter a percentage of your choice via the navigation buttons, see chpt.
14.2.3 Percent determination or In weighing mode press MENU for 3 sec., the currently set percentage determination mode is displayed. Place an empty container on the balance and tare it by pressing TARE. Place goods to be weighed on balance. The weight of the sample is displayed in percentage in terms of the reference weight. Either carry out more percent weighing or The menu item to change the stored reference appears. + see chpt. 14.2.2 / 14.2.
14.3 Formula mode The formula function allows to add on various components of a mixture. For check purpose, the weight of all components (CMP001, (CMP002 etc.), as well as the total weight (TOTAL) can be printed out. The balance works with a separated memory for the weight of the weighing container and of the formula components. In formula mode the „Auto zero“ function is not active (see chpt. 12.2). 1. Connect printer (see chpt. 15 „Data output“) 2.
Weigh further components in the same way During formulation weighing can be displayed at any time by pressing MENU (for 3 sec.). 5. Complete formulation process Press ON/OFF. When the symbol [G] is displayed, the total weight (TOTAL) of all components is displayed and edited to the printer. When the symbol is displayed, the balance is ready for further measurements. 6. Return to weighing mode Press ON/OFF for 3 sec. Printout example „GLP ON“ (KERN YKB-01N): Company Model Serial no. ID no.
14.3.1 Enable edition „component no.“ Select formula mode, see chpt. 14.3 When the symbol button twice. appears, press the MENU Press the navigation buttons ( ) repeatedly until „ELM.NUM“ is displayed. Confirm with PRINT. The current setting is marked by the stability display ( ). with ( ) without ( ) Edition „component no.“ (e.g. CMP001) No edition „component no.“ Change using TARE. Back to formula mode Press ON/OFF repeatedly or for 3 sec.
14.3.2 Enable issue „TOTAL“ Select formula mode, see chpt. 14.3 When the symbol button twice. appears, press the MENU Press the navigation buttons ( ) repeatedly until „TOTAL“ is displayed. Confirm with PRINT. The current setting is marked by the stability display ( ). with ( ) without ( ) Issue „TOTAL“ No issue „TOTAL“ Change using TARE. Back to formula mode Press ON/OFF repeatedly or for 3 sec.
15 Check weighing and target mode (check and target weighing) • The check weighing or target mode can be applied for the application functions (see chpt. 14). • The balance starts in the mode, in which it has been switched off. 15.1 Check weighing In many cases not the nominal value of the weighed goods is the decisive parameter, but the deviation from this nominal value. Such applications are for example the weight check of equivalent packages or the process check of parts in a fabrication process.
0. Call up menu In weighing mode press the MENU button twice 1. Select function Press the navigation buttons ( ) repeatedly until „TOOLS“ is displayed. Confirm with PRINT Press the navigation buttons ( ) repeatedly until „CHECK.W“ is displayed. The current setting is marked by the stability display ( ). Further steps depend on the desired setting: Stability display ( yes ) Function set/update cancel ON Press PRINT and continue with step 3. Press TARE and continue with step 4.
Confirm with TARE. Press the navigation buttons ( ) repeatedly until „LO.LIM“ is displayed. Press TARE. The current setting will be displayed. Use the navigation buttons ( ) to enter the desired value, see chpt. 3.1.1 „Numeric input“. Confirm with TARE. Press the navigation buttons ( ) repeatedly until „UND.RNG“ is displayed. Press TARE. The current setting will be displayed. Use the navigation buttons ( ) to enter the desired value, see chpt. 3.1.1 „Numeric input“.
3. Return to weighing mode Press ON/OFF repeatedly or for 3 sec. 4. Start tolerance control If necessary, place an empty container on the balance and tare it by pressing TARE. or appears. With Place weighed goods and wait until the indicator , the help of the indicator check if the weighed goods are under, inside or over the default tolerance Example of entry: Sample weight < UND.RG (sample weight < 5 .0000g) UND.RG ≤ sample weight < LO.LIM (sample weight 5.0000g - 5.9999g) LO.LIM ≤ sample weight ≤ HI.
15.2 Target mode This mode e.g. is used for weighing constant liquid quantities or for assessment of missing quantities or excess quantities. The target value is the numeric value which corresponds to the nominal quantity of the used unit. Beside the target value a tolerance value is entered. This is a numerical value which is plus/minus over or under the acceptable target value. The indicators or advise, when the target value is reached.
1. Call up menu In weighing mode press the MENU button twice 2. Select function Press the navigation buttons ( ) repeatedly until „TOOLS“ is displayed. Confirm with PRINT Press the navigation buttons ( ) repeatedly until „TARGT“ is displayed. The current setting is marked by the stability display ( ). Further steps depend on the desired setting: Stability display ( ) yes no Function set/update cancel ON Press PRINT and continue with step 3. Press TARE and continue with step 4.
Confirm with TARE. 4. Return to weighing mode Press ON/OFF repeatedly or for 3 sec. 5. Start tolerance control If necessary, place an empty container on the balance and tare it by pressing TARE. Place weighed goods and wait until the indicator or appears. With , the help of the indicator check if the weighed goods are under, inside or over the default tolerance Example of entry: Weight less than nominal weight and below the lower tolerance TG.VAL 100.0000 g LM.VAL 10.
16 Data output The appliance interface allows a bi-directional data exchange from the balance to external devices. This data exchange is asynchronous using ASCII - Code. The following conditions must be met to provide successful communication between the weighing balance and the printer. • Use a suitable cable to connect the weighing balance to the interface of the printer/PC. Faultless operation requires an adequate KERN interface cable (optional).
16.2 Issue functions 16.2.1 Automatic data output / Auto Print function Data output ensues automatically without pressing PRINT, as soon as the respective output condition is fulfilled. This is defined by the menu setting. Tab. 1: stable / positive Mode 1 LD. . Mode 2 LD.UL. Mode 3 LD. .Z stable / negative - Stabilisation / zero display Check weighing - Output for stable and positive weighing value - Output for stable and positive or negative weighing value.
Further steps depend on the desired setting: Stability display ( ) yes no Function set/update ON Press PRINT and continue with step 3. off Press TARE and continue with step 3. cancel Press TARE and continue with step 5. continue with step 5. 3. Set output condition Use the navigation buttons ( ) to select the desired mode, e.g. Mode 3 (details see Tab. 1) Confirm with TARE 4.
16.2.2 Continuous Output function 1. Invoke Data Output menu Press PRINT for 3 sec. in weighing mode. 2. Select function Press the navigation buttons ( ) repeatedly until „APL.PRN“ is displayed. Confirm with PRINT Press the navigation buttons ( ) repeatedly until „SEQ.PRN“ is displayed. The current setting is marked by the stability display ( ). with ( ) on without ( ) off Change using TARE. 3.
4. Connect filter Press the navigation buttons ( ) repeatedly until „NO.FIL“ is displayed. The current setting is marked by the stability display ( ). with ( ) without ( ) Weighing value is filtered Weighing value is not filtered Change using TARE 5. Return to weighing mode Press ON/OFF repeatedly or for 3 s. With activated menu setting „MANU“ (see step 3, with the indicator will appear. 6. 7. 8.
16.2.3 Output Timing Change function This function allows to select if the data output shall ensue with stable or instable weighing value after pressing PRINT. 1. Invoke Data Output menu Press PRINT for 3 sec. in weighing mode. 2. Select function Press the navigation buttons ( ) repeatedly until „APL.PRN“ is displayed. Confirm with PRINT Press the navigation buttons ( ) repeatedly until „PRINT.F“ is displayed. Confirm with PRINT. The current setting is marked by the stability display ( ).
16.2.4 GLP Output function / balance identification number With the GLP Output function the printouts of weighing results are completed with a bottom row and a head line. Contents of the headline and bottom row see following printout example. + + Enable GLP Output function, see chpt. 8.3 Enter balance identification number, see chpt. 8.4 Printout example: KERN & Sohn GmbH TYPE ABJ 220-4NM SN WBIIAB000I ID 1234 200.0000g -SIGNATURE- Company Model Serial no. Balance identification no. (see chpt. 8.
16.3 Communication parameters By calling-up a standard setting „MODE 1 - MODE 5“ all communication parameters are already preset (see chpt. 16.3.1). The subsequent standard setting must be selected according to the printer (details see the following table). In the menu item „MODE U“ all parameters can be set defined by user (see chpt. 16.3.2).
16.3.1 Selection of a standard setting „MODE 1 - MODE 5“ 1. Invoke Data Output menu Press PRINT for 3 sec. in weighing mode. 2. Select function Press the navigation buttons ( ) repeatedly until „COMM.SET“ is displayed. Confirm with PRINT Press the navigation buttons ( ) repeatedly until the desired mode „MODE 1 - MODE 5“ is displayed. The current setting is marked by the stability display ( ). with ( ) on without ( ) off Change using TARE. 3.
16.3.2 User-defined settings „MODE U“ / setting for KERN –YKB-01N In the menu item „MODE U“ every communication parameter can be set individually. 1. Invoke Data Output menu Press PRINT for 3 sec. in weighing mode. 2. Select function Press the navigation buttons ( ) repeatedly until „COMM.SET“ is displayed. Confirm with PRINT Press the navigation buttons ( ) repeatedly until the desired mode „MODE 1 - MODE 5“ is displayed. The current setting is marked by the stability display ( ).
3. Set the communication speed (Baud rate) Press PRINT Press the navigation buttons ( ) repeatedly until „START“ is displayed. Press PRINT. The current setting is marked by the stability display ( ). Press the navigation buttons ( ) repeatedly until the desired setting (e.g. 9600 bps) appears. Press TARE. Selectable settings: Display B0.300 B0.600 B0.1200 B0.2400 B0.4800 B0.9600 Baud rate 300bps 600bps 1200bps 2400bps 4800bps 9600bps B.19.2k B.38.4k 19.2k bps 38.4k bps 4.
5. Stop bit settings Use ON/OFF to return to menu Press the navigation buttons ( ) repeatedly until „STOP“ is displayed. Press PRINT. The current setting is marked by the stability display ( ). Press the navigation buttons ( ) repeatedly until the desired setting (e.g. S. 1) appears. Press TARE. Selectable settings: Display S. 1 S. 2 Stop bit 1 bit 2 bit 6. Set handshake Use ON/OFF to return to menu Press the navigation buttons ( ) repeatedly until „HAND.SHK“ is displayed.
7. Set data format Use ON/OFF to return to menu Press the navigation buttons ( ) repeatedly until „D.FORM“ is displayed. Press PRINT. The current setting is marked by the stability display ( ). Press the navigation buttons ( ) repeatedly until the desired setting (e.g. DF.1) appears. Press TARE. Selectable settings (details see chpt. 15.4): Display DF.1 DF.2 DF.3 DF.4 DF.
16.4 Data Formats In the menu item „D.FORM“ four data formats „DF.1-DF.4“ are available. + Menu setting, see chpt. 16.3.2, step 7 „Setting data format“. 1. Data format 1 „DF.1“ Example 9.9949 g: Position data ASCII code Explanation 1 20H Positive weighing value = space character 20H Negative weighing value = minus sign 2DH 2 20H 3 20H 4 9 39H 5 . 2EH 6 9 39H 7 9 39H 8 4 34H 9 9 39H 10 g 67H 11 12 Numeric weighing value is displayed in 8 positions.
2. Data format 2 „DF.2“ Example 9.9949 g: Position Data ASCII code Explanation 1 20H Positive weighing value = space character 20H 2 20H Negative weighing value = minus sign 2DH 3 U 55H 4 20H 5 20H 6 20H 7 20H 8 20H 9 9 39H 10 . 2EH 11 9 39H 12 9 39H 13 4 34H 14 9 39H 15 20H 16 g 67H 17 C/R 0DH Stable weighing value = S (stable) 53H Unstable weighing value = U (unstable) 55H Numeric weighing value is displayed in 8 positions.
3. Data format 3 „DF.3“ Example 9.9949 g: Position Data 1 + ASCII code Explanation 2BH Positive weighing value = plus sign 2BH 2 20H Negative weighing value = minus sign 2DH 3 20H 4 20H Numeric weighing value is displayed in 8 positions. 5 9 39H 6 .
4. Data format 4 „DF.4“ Example 9.9949 g: Position Data 1 S ASCII code Explanation 53H Stable weighing value = S (stable) 53H 2 20H Unstable weighing value = U (unstable) 55H 3 20H 4 2BH 5 20H 6 + 2BH 7 9 39H 8 . 2EH 9 9 39H 10 9 39H 11 4 34H 12 9 39H 13 20H 14 20H 15 g 67H 16 C/R 0DH Positive weighing value = plus sign 2BH Negative weighing value = minus sign 2DH Numeric weighing value is displayed in 8 positions.
16.5 Remote control instructions Command Function D01 Continuous data output D02 Continuous data output of stable weighing values D03 Status of stability display is attached to the data in the continuous output. U: instable S: stable D05 Single output D06 Automatic output D07 Single output. Status of stability display is attached to the data in the output. U instable S: stable D08 Single output with stable weighing value D09 Cancel output BREAK Q CAL Function as , see chpt. 3.
17 Service, maintenance, disposal 17.1 Cleaning Before cleaning, disconnect the appliance from the operating voltage. Fig. 1: Clean the balance 1. Display Please do not use aggressive cleaning agents (solvents or similar agents), but a cloth dampened with mild soap suds. 2. Weighing plate Remove weighing plate, clean it wet and dry it before installation 3. casing, housing, case, box Please do not use aggressive cleaning agents (solvents or similar agents), but a cloth dampened with mild soap suds.
1. Remove, screening ring, weighing plate and carrier of weighing plate 2. Remove the plastic handle by turning. Do no touch the support of the weighing plate. This could cause damage to the balance. 3. Remove glass door carefully acc. to fig. Fig. 2: Remove the glass doors 4. Re-install the glass door in reverse order. To secure the glass door always reattach the plastic handle.
17.2 Service, maintenance The appliance may only be opened by trained service technicians who are authorized by KERN. Before opening, disconnect from power supply. 17.3 Disposal Disposal of packaging and appliance must be carried out by operator according to valid national or regional law of the location where the appliance is used. 18 Instant help Possible causes of errors: In case of an error in the program process, briefly turn off the balance and disconnect from power supply.
18.1 Error messages Error message Explication ERR H Hardware error High zero point shift during adjustment ERR C Objects present on weighing plate. Remedy Switch balance off and on again. If the error message remains inform manufacturer. Use ON/OFF to return into weighing mode. Restart adjustment process. Missing weighing plate CAL D Display instable Check ambient conditions (draught, vibrations etc.) Use ON/OFF to return into weighing mode. Restart adjustment process.