Instruction Manual

ABT-BA-e-1312
62
Printout:
---ADDON Mode---
CMP001=
5,0003 g
CMP002=
5,0017 g
CMP003=
5,0010 g
TOTAL=
15,0030 g
Note: Only one function at a time may be activated (on) when working with func-
tions auto print, add-on and recipe.
Whilst the function is active, automatic adjustment (PSC/Clock-CAL) does
not take place. A flashing symbol, however, indicates that adjustment is
required.
14.6 Recipe mode
The formula function allows to add on various components of a mixture. The total
weight of all components can be invoked to check (separate memory for the weight
of tare vessel and recipe components).
Activate function "FormU: on“ (see chap. 14.5 Add up):
Whilst the recipe function is active, the display keeps showing the "FormU -on" sym-
bol [ M].
Operation Display
When using a container, place this on the weighing plate
and tare.
0.0000
g
Weigh first component.
0.5361
g
When standstill control ( ) is complete, the weighing
value of the first component (CMP001) is issued to the
optional printer by pressing the [PRINT] key. The dis-
played value is added into the total adding memory. Af-
terwards automatic taring will take place.
0.5361
g
0.0000
g
Weigh second component
0.5422
g
When standstill control ( ) is complete, the weighing
value of the second component (CMP002) is issued to
the optional printer by pressing the [PRINT] key. The
displayed value is added into the total adding memory.
Afterwards automatic taring will take place.
0.54221
g
0.0000
g