Owner manual

168 AET_/PET_/ILT_NM-BA-e-1410
Switch on continuous transmission in current measuring unit
Format: CU1 CR LF
Accessible responses:
CU1_I CR LF
- command understood but not accessible at this moment
CU1_A CR LF
MASS FRAME
- command understood and in progress
- response is mass value in current measuring unit
Frame format:
1
2
3
4
5
6
7-15
16
17
18
19
20
21
S U I
stability
marker
space character mass space unit CR LF
Switch off continuous transmission in current weighing unit
Format: CU0 CR LF
Accessible responses:
CU0_I CR LF
- command understood but not accessible at this moment
CU0_A CR LF
- command understood and in progress
Set low checkweighing threshold LO
Format: DH_XXXXX CR LF, where XXXXXmass format
Accessible responses:
DH_OK CR LF
- command carried out
ES CR LF
- command not recognized (incorrect mass format)
Set high checkweighing threshold HI
Format: UH_XXXXX CR LF, where XXXXXmass format
Accessible responses:
UH_OK CR LF
- command carried out
ES CR LF
- command not recognized (incorrect mass format)
Give value of low checkweighing threshold LO
Format: ODH CR LF
Accessible response: DH_MASS CR LF command carried out
Frame format:
1 2 3 4-12 13 14 15 16 17 18 19
D H space mass space unit space CR LF
Mass - 9 characters with right justification
Unit - 3 characters with right justification