User Manual
ALT_B-PLT_A-PLT_F-BA-e-1333    29 
8.1  Automatic adjustment with internal weight 
The automatic adjustment with internal weight is started automatically, if the balance 
  after the weighing balance was disconnected from the mains 
  after pressing ON/OFF in stand-by mode 
  after a temperature change of 1.5 °C  
with non loaded weighing pan / zero display 
  after a time interval of 20 min with non loaded weighing pan / zero 
display 
The automatic adjustment function is always enabled. At any time, however, you can 
carry out a manual adjustment (press CAL-button) using the internal adjustment 
weight, see chap. 7.1. 
Or selecting the adjustment mode in the menu: 
  In weighing mode press the MENU button and keep it pressed until the acoustic 
signal gets mute. The setup menu is displayed. 
  Use the navigation buttons  to select the menu item „Adjustment mode“. 
  Confirm by the PRINT button, „Autom. Just.“ will be displayed. 
  Take over selection using the PRINT button. 
Weighing balance returns to menu. Either make more settings in the menu or go 
back to weighing mode as follows. 
  Press the MENU button and keep it pressed until the acoustic signal gets mute. 
The balance returns automatically into weighing mode. 