12.1  Parts counting 
Before the balance can count parts, it must know the average part weight (i.e. 
reference). Proceed by putting on a certain number of the parts to be counted. The 
balance determines the total weight and divides it by the number of parts, the so-
called reference quantity. Counting is then carried out on the basis of the calculated 
average piece weight. 
As a rule: 
The higher the reference quantity the higher the counting exactness. 
12.1.1  Determine reference by weighing 
  In weighing mode press MENU button. The master menu will be displayed. 
  Use the navigation buttons  to select the menu item „Piece counting“. 
  Acknowledge using PRINT button, the currently set reference quantity is 
displayed. 
  Use the navigation buttons  to select the desired setting. 
Display 
Description 
10 off 
Reference piece number 10 
25 off 
Reference piece number 25 
50 off 
Reference piece number 50 
100 off 
Reference piece number 100 
Manual 
Numerical input of reference weight, see chap. 12.1.2 