User Manual
82      ALT_B-PLT_A-PLT_F-BA-e-1333 
12.5  Percent determination 
Percent determination allows weight display in percent, in relation to a reference 
weight. 
12.5.1  Entering the reference weight by weighing 
  In weighing mode press MENU button. The master menu will be displayed. 
  Use the navigation buttons  to select the menu item „Percent weigh.“. 
  Confirm with the PRINT button, the currently set mode will be displayed. 
  Use the navigation buttons  to select setting „Automatic“. 
  If using a weighing container, put this before pressing the PRINT button. The 
taring process is automatically started. 
  Press the PRINT key. 
  Place the reference weight (=100 %) and press the PRINT button. As soon the 
weighing value is stable, the determined weight will be taken over as reference 
(100%). From now the balance is in percent determining mode. 
  Place weighed goods, the result display will appear. 
W% 
Weight of the sample in percentage 
W 
Weight of the sample in grams 
REF: 
Reference weight (100%) 
When an optional printer is connected, the displayed value can be edited 
using PRINT. Printout example (KERN YKB-01N): 
07-09-11 11:14:57 
Perc. 
49.95 % 
Weight of the sample in 
percentage 
Weight: 
9.990 g 
Weight of the sample in grams 
Refer.: 
20.000 g 
Reference weight (100%) 