Manual
KFP V20_600-6000-IA-e-1313      17 
9.2  Check and adjustment of the corner load 
Check of the corner load: 
  Place the test weights in the centre of the load plate 
and tare. 
  The balance displays -0-. 
  Place the test weights successively on all four 
corners. 
  Now the deviations are displayed with sign, write 
down the values. If there are deviations out of the 
tolerances (see chap. 9.1), an adjustment will be 
necessary. 
Adjustment of corner load: 
Preparation: 
  For a better control of the modifications which occur 
during adjustment, select in the configuration menu 
the highest readability for control purposes. 
  Open connection box 
Adjustment rule: 
The corner (weighing cell) with the biggest negative 
deviation must be set to zero. Do not re-adjust this corner 
even after several adjustment sequences. 