Manual
34        CFS/CCS-BA-e-1220 
9  Basic Operation 
9.1  Switching on/off 
  to switch-on actuate forward the switch-on/switch-out on the right lower side of 
the balance. The balance will carry out a self-test As soon as the weight display 
appears, the balance is ready for weighing. 
  To switch off press the on/off switch on the right bottom side of the balance at 
the back in the CFS 50K-3 models also press forwards. 
9.2  Zeroing 
Resetting to zero corrects the influence of light soiling on the weighing plate. The 
resetting range of the balance was adjusted to ± 2% max. by factory. 
Further adjustments are possible in the menu, see chapter 13. 
When using as counting system the zeroing range of both balances can be set in 
the menu, see chapter 13. 
Manual 
  Unload the balance 
  Press the ZERO button, the balance begins to reset to zero. 
The [] symbol over a appears. 
Automatic 
In the menu the automatic zero point correction can be switched off or the amount 
can be changed, see chapter 13. 
9.3  Switch over balance/weighing bridge 
For piece counting a weighing bridge can be connected via the second balance 
interface. In the counting system KERN CCS the quantity counting is made on the 
bulk material balance KERN KFP. The reference balance KERN CFS permits due 
to its high resolution a very precise determination of the average piece weight. 
The second balance can be operated in the same manner like the first balance. 