KERN & Sohn GmbH Ziegelei 1 D-72336 Balingen E-Mail: info@kern-sohn.com Phone: +49-[0]7433- 9933-0 Fax: +49-[0]7433-9933-149 Internet: www.kern-sohn.com Operating manual Counting balance/counting system KERN CFS/CCS Version 2.
GB KERN CFS/CCS Version 2.2 11/2014 Operating instructions Counting balance/counting system Contents 1 Technical data .......................................................................................... 4 1.1 1.2 KERN CFS .......................................................................................................................... 4 Counting systems KERN CCS .......................................................................................... 7 2 Appliance overview ............
8.1 8.2 8.3 8.4 Determination of the average piece weight by weighing ............................................. 28 Numeric input of the average piece weight ................................................................... 31 Automatic reference optimization .................................................................................. 33 Counting with counting system...................................................................................... 34 9 Fill-to-target function .............
1 Technical data 1.1 KERN CFS KERN Readability (d) Weighing range (max) Reproducibility Linearity CFS 300-3 CFS 3K-5 CFS 6K0.1 0.001 g 0.01 g 0.1 g 300 g 3 kg 6 kg 0.002 g 0.02 g 0.1 g ± 0.004 g ± 0.04 g ± 0.
KERN CFS 15K0.2 CFS 30K0.5 CFS 50K-3 Readability (d) 0.2 g 0.5 g 1g Weighing range (max) 15 kg 30 kg 50 kg Reproducibility 0.2 g 0.5 g 1g ± 0.
Dimensions: Models • • • • • CFS 300-3 CFS 3K-5 CFS 6K0.1 CFS 15K0.2 CFS 30K0.
1.2 Counting systems KERN CCS Model Bulk material balance KERN KFP Weighing range Readab ility [Max] kg [d] g Weighing pan Recommended calibration weight, not Reference scale included, kg [class F1] CFS Weighing range Readability [Max] [d] g g Minimum piece weight [count] g/piece CCS 6K-6 KFP 6V20M 6 2 230x230x100 6 CFS300-3 300 0.001 0.005 CCS 10K-6 KFP 15V20M 15 5 300x240x100 15 CFS 300-3 300 0.001 0.005 CCS 30K0.
Model Bulk material balance KERN KFP Weighing range Reada bility [Max] kg [d] g Weighing pan Recommended calibration weight, not Reference scale included, kg [class F1] CFS Weighing range Readability [Max] [d] g g Minimum piece weight [count] g/piece CCS 600K-2 KFP 600V20SM 600 0.01 1000x1000x80 600 CFS 3K-5 3000 0.01 0.05 CCS 600K-2L KFP 600V20M 600 0.01 1500x1250x80 600 CFS 3K-5 3000 0.01 0.05 CCS 600K-2U KFU 600V20M 600 0.01 840x11190x90 600 CFS 3K-5 3000 0.
2 Appliance overview 2.1 Counting scales KERN CFS Model: CFS 300-3 Models: CFS 3K-5, CFS 6K0.1, CFS 15K0.2, CFS 30K0.5 1. Weighing plate / rechargeable battery compartment (under weighing plate) 2. Windshield 3. Bubble level 4. RS 232 interface 5. Second balance interface 6. Footscrews 7. ON/OFF switch 8.
Model CFS 50K-3 1. 2. 3. 4. 5. 6. 7.
2.2 Counting systems KERN CCS In factory the counting system KERN CCS is preconfigured in a way that no more changes will be necessary. Bulk material balance KERN KFP 2.3 Reference balance KERN CFS Counting systems with your selected bulk scales When attaching a bulk scales (not pre-adjusted by KERN), the following must be complied with: Connect bulk scales with a suitable cable via the second-balance interface. Interface connection allocation cf. chapter 16 Adjustment of bulk scales, cf.
Example 2: High-load reference scale Bulk material balance KERN KFP 12 Reference scale KERN CFS 50K-3 CFS/CCS-BA-e-1422
2.4 Overview of display Models CFS 300-3, CFS 3K-5, CFS 6K0.1, CFS 15K0.2, CFS 30K0.5: Weight Average piece weight Quantity Accumulator’s loading state, cf. chapter 6.5 Model CFS 50K-3: Weight Accumulator’s loading state, cf. chapter 6.
2.4.1 Display weight Here the weight of your goods is displayed in [kg]. Indicator [] atop the symbol shows: Storage battery status display NET Net weight Stability state indication Model CFS 50K-3 a Zeroing display lb/kg Current weighing unit 2.4.2 Display average piece weight Here the average reference weight of a sample is displayed in [g]. This value is either numerically entered by user or calculated by weighing on balance.
2.5 Keyboard overview Models CFS 300-3, CFS 3K-5, CFS 6K0.1, CFS 15K0.2, CFS 30K0.5 Selection Function in Weighing mode • Numeric keys • • Decimal point At numeric input digit selection to the left • Delete • • • • Totalization Display total weight /number of weighings / total quantity At numeric input digit selection to the right Data output (menu setting , see chapter 12.2) • Store/call article, cf. chapters 11.1 / 11.2 • Fill-to-target function (cf.
• • Taring Confirm • • Zeroing Back to menu/weighing mode Model CFS 50K-3: Selection Function in Weighing mode • Article direct keys, cf. chapter 11.3 • Numeric keys • Decimal point • Delete ….. …..
• • • Summing up / printing (menu adjustment , cf. chapter 12.2 Display total weight /number of weighings / total quantity Data output (menu setting , see chapter 12.2) • Fill-to-target function (cf. chapter 9) • Store/call article, cf. chapters 11.1 / 11.2 • • Switching over the scale, cf. chapter 7.3 At numeric input digit selection to the left • • Input of the average piece weight by weighing, see chap. 8.1 Leafing through the menu • • Numeric input of the average piece weight see chapter 8.
3 Basic instructions 3.1 Proper use The balance / counting system you purchased is intended to determine the weighing value of material to be weighed. It is intended to be used as a “nonautomatic balance”, i.e. the material to be weighed is manually and carefully placed in the centre of the weighing pan.. As soon as a stable weighing value is reached the weighing value can be read. 3.2 Improper Use Do not use balance / counting system for dynamic weighing.
3.4 Monitoring of Test Resources In the framework of quality assurance the measuring-related properties of the balance and, if applicable, the testing weight, must be checked regularly. The responsible user must define a suitable interval as well as type and scope of this test. Information is available on KERN’s home page (www.kern-sohn.com with regard to the monitoring of balance test substances and the test weights required for this.
6 Unpacking, Setup and Commissioning 6.1 Installation Site, Location of Use The balances / counting systems are designed in a way that reliable weighing results are achieved in common conditions of use. You will work accurately and fast, if you select the right location for your balance / counting system.
6.3 Installing / removing transport fittings Remove transport lock, if applicable. KERN CFS 3K0.5, CFS 6K0.1: KERN CFS 300-3: Bulk scales KERN KFP (illustration example): KERN KFP 6V20M, KFP 6V20LM, KFP 15V20M. Please take further details from the installation manual, added to the platform.
Install weighing plate and wind screen, if applicable, if needed. Level balance with foot screws until the air bubble of the water balance is in the prescribed circle. Check levelling regularly Interconnect reference scale and bulk scales via the second-scale interface in the counting systems KERN CCS. 6.4 Mains connection Power is supplied via the external mains adapter. The stated voltage value must be the same as the local voltage. Only use original KERN mains adapters.
6.6 Connection of peripheral devices Before connecting or disconnecting of additional devices (printer, PC) to the data interface, always disconnect the balance from the power supply. With your balance, only use accessories and peripheral devices by KERN, as they are ideally tuned to your balance. 6.7 Initial Commissioning In order to obtain exact results with the electronic balances, your balance must have reached the operating temperature (see warming up time chap. 1).
7 Basic Operation 7.1 Switching on/off Actuate power switch (cf. chapter 2) at the right bottom of the scale to the front. The balance will carry out a self-test As soon as the weight display appears, the balance is ready for weighing. To switch-off push backward the switch-on/switch-out on the right lower side of the balance. 7.2 Zeroing Resetting to zero corrects the influence of light soiling on the weighing plate. The resetting range of the balance was adjusted to ± 2% max. by factory.
Display example model CFS 6K0.1: (1) Reference balance KERN CFS (2) Bulk scales, e.g.
7.4 Weighing with tare A tare value can be entered for the reference as well as for the bulk material balance. Before setting a tare value select active balance, see chap. 9.3. 7.4.1 Taring Deposit weighing vessel. After successful stop check press the TARE button. Zero display and the indicator [] above NET appear. The weight of the container is now internally saved. Weigh the material, the net weight will be indicated.
8 Parts counting Before the balance can count parts, it must know the average part weight (i.e. reference). Proceed by putting on a certain number of the parts to be counted. The balance determines the total weight and divides it by the number of parts, the socalled reference quantity. Counting is then carried out on the basis of the calculated average piece weight. As a rule: The higher the reference quantity the higher the counting exactness.
8.1 Determination of the average piece weight by weighing Set reference Reset balance to zero or tare the empty weighing container if necessary. Place on the weighing plate a known number (e.g. 10 items) of individual pieces as a reference. Input the number of reference pieces via the number keys. Wait for stability display and confirm with or (model CFS 50K-3) within 5 sec. The scale shall detect the average piece weight and will then display the number of pieces. Display example model CFS 6K0.
Display example model CFS 50K-3: CFS/CCS-BA-e-1422 29
Count the items Tare if necessary, place weighing good and read off the number of items. Display example model CFS 6K0.1: Display example model CFS 50K-3: Interconnecting an optional printer, the display value will be output by pressing (menu adjustments F1 oFF ACC off; F2 Prt P mode Print Au OFF, cf. chapter 12.2). Printout example KERN YKB 01N / CFS 6K0.1: S1 Active balance, see chap. 7.3 ID: 123456 N 2.4986 kg 49.9755 g/pcs 50 pcs User identification number (cf. chapter 12.
8.2 Numeric input of the average piece weight Set reference Input known average piece weight, e.g. 10 g, with the numeric keys and confirm with or (models CFS 50K-3) within 5 sec. If in the weight display as weighing unit [kg] is active, the average piece weight will be displayed in [g]. If as weighing unit [lb] is active, the average piece weight is also displayed in [lb]. Display example model CFS 6K0.
Display example model CFS 50K-3: Count the items Tare if necessary, place weighing good and read off the number of items. Interconnecting an optional printer, the display value will be output by pressing , display and printout examples cf. chapter 10.1.
8.3 Automatic reference optimization If at the reference determination the placed weight or the placed piece number is too small, in the display of the average piece weight the triangle symbol will appear over [ ] or [ ]. To optimize the calculated average piece weight automatically, add further parts whose number is smaller than that of the first reference determination. After the reference optimization sounds a signal tone. For each reference optimization the average piece weight is newly calculated.
8.4 Counting with counting system (Illustration example) Bulk material balance e.g. KERN KFP Reference balance KERN CFS • Here pieces of huge quantities will be counted. • Due to its high resolution it is useful for accurate determination of the average piece weight. • Big parts (max > 3kg) are counted on the platform. • Smallest parts (max < 3kg) are counted on the precise KERN CFS.
Display example model CFS 6K0.1: In order to avoid errors at the piece number determination, both balances must be adjusted with the same acceleration due to gravity (see chap. 14).
9 Fill-to-target function The balance allows weighing of goods to a certain target weight or target piece number within defined tolerances. With this function one can also check if the weighing good is within a defined tolerance range. Tolerance control is possible in weighing or target mode. Reaching the target value is indicated by an acoustic (if activated in menu) and an optic signal (tolerance mark ). Audio signal: The acoustic signal is dependent of the adjustment in the menu block „F1 oFFBEEP“.
9.1 Tolerance check for target weight Press , the active tolerance weighing mode is indicated. If needed, select tolerance control for target weight (PSt nEt) with (models CFS 50K-3). or Display example model CFS 6K0.1: Press the TARE key , which shows the currently set upper threshold. To change, input desired value, e.g. 5.500 kg with the numeric keys. Confirm with TARE key; the currently adjusted minimum threshold will be displayed. To change, input desired value, e.g.
Put on the weighed material and check whether the weighed material is within the given tolerance, alongside the tolerance mark / acoustic signal. Display of tolerance mark , if weighing goods are under the specified tolerance: Display of tolerance mark , if weighing goods within the specified tolerance: Display of tolerance mark , if weighing goods above the specified tolerance: For tolerance control, also only one limit value can be set.
9.2 Tolerance check for target quantity Press , the active tolerance weighing mode is indicated. If needed, select tolerance control for target weight (PSt nEt) with (models CFS 50K-3). or Display example model CFS 6K0.1: Press the TARE key , which shows the currently set upper threshold. To change, input the desired value, e.g. 100 pieces, with the numeric keys. Confirm with TARE key; the currently adjusted minimum threshold will be displayed. To change, input the desired value, e.g.
Determine the average piece weight (see chap. 10.1 or 10.2), place the weighing goods and check using the tolerance mark , if the number of the placed pieces is under, within or above the specified tolerance.
10 Totalization Tolerance control is possible in weighing or target mode. When using in the counting system, no matter, if the weighed goods are on the reference or on the bulk material balance. Preparing: When using as counting system select using the balance, on which will be totalized. The appeared [] displays the current balance. When totalizing in counting mode set the average piece weight. (see chap. 8.1 or 8.2) If necessary, tare the empty balance container. 10.
This process can be repeated 99 times or until the weighing range of the balance is exhausted. Display of the saved weighing data: Press , the total weight, weighing number, and total number of pieces will be displayed and be output if connected to an optional printer. Display example model CFS 6K0.1: Total weight placed on balance: Number weighing processes. Total number of pieces: Printout example KERN YKB-01N: S1 ID: Active balance, see chap. 7.3 123456 User identification number (cf.
10.2 Automatic adding-up This function is used to add up the individual weighing values automatically into the sum storage once the scale is relieved, without pressing 50K-3); and edited if interconnected to an optional printer. • or (models CFS Menu settings: „F1 off“ „ACC“ „ON“ (unavailable in model CFS 50K-3) „F2 Prt“ „P mode“ „Print“ „Au ON“, cf. chapter 12.2 • When using as counting system there can be totalized on the reference as well as on the bulk material balance.
11 Store article information The scale has more than 100 article storage places for frequently used tare values, average piece weights and article names. These data can be accessed by calling up the corresponding number of a certain article. Model CFS 50K-3 is additionally provided with 5 direct keys 11.3). ~ , cf. chapter 11.1 Store article Preparation: If necessary set balance to zero using . Tare if using a weighing container.
Store article: Press to input the storage place number (e.g. number 27) Display example model CFS 6K0.1: Input „2“ and „7“ with the numeric keys. , the currently stored article name will be displayed. The first digit is Press flashing. Delete by , if necessary, and overwrite with new article name as described below (no more than 12 characters, e.g. „KERN 1234 AB“).
For input of numbers, actuate shortly the numeric button. For input of letters press the numeric button and keep it pressed until the desired letter is displayed. The characters according to keyboard assignation run through. 1 2 3 4 5 6 7 8 9 0 -/\ ABC DEF GHI JKL MNO PQRS TUV WXYZ _ [] _ = Space Overview data input / data output: Use flashes. to move number selection to the left, the respective active position to move the number selection to the right, the respective active Use position flashes.
11.2 Call article to select the balance, on which the When using as counting system use tare value is stored. The appeared [] displays the current balance. Press ; „PLU“ to input the storage place number will show. Call desired number, e.g. 27; press numeric keys „2“ and „7“ to this end. again, the storage place number (e.g. PLU 27) and the article name Press will be shown for 1 second. If the data shall be displayed longer time, keep pressed.
When connecting an optional printer, the data can be edited by pressing Printout example KERN YKB-01N: S1 ID: Active balance, see chap. 7.3 123456 KERN 1244 AB N. 1.9990 kg 10 g/pcs 200 pcs User identification number (cf. chapter 12.2) Article name (cf. chapter 11.1) Placed net weight Average piece weight Quantity placed on balance Further printout examples cf. chapter 17.2. 11.3 Article direct keys ~ (only model CFS 50K-3) 1. Preparation, cf. chapter 11.1 2. Store article for approx.
3. Call article Press direct key e.g. be shown for 1 second. ; the storage place number and the article name will The display changes into the counting mode, the stored tare value e.g. 500 g and the average piece weight e.g. 10g /pce. are displayed. Place load on pan and read the number of pieces. If connected to an optional printer, the data will be added up into the sum storage and edited by pressing . Printout example CFS 50K-3 / KERN YKB 01N: LOCAL SCALE Active balance, see chap. 7.
12 Menu The menu is structured in the following menu blocks. 1. Balance settings 2. Settings serial interface 3. Input/display user identification number 4. Input/display scale identification number 5. Configuration of bulk scale 12.1 Navigation in the menu Call up menu Select menu block Switch-on balance and during the selftest press The first menu block is displayed. or (model CFS 50K-3) the separate menu With items can be accessed in succession. Select menu item .
12.2 Menu overview 12.2.1 Models 6K0.1, 15K0.2, 30K0.5, 50K-3 Menu block Menu item Available Main menu Submenu Settings Explanation Signal tone switched off Signal tone on, if weighing value within tolerance limits Signal tone on, if weighing value outside tolerance limits Display background illumination off resp.
Baud rate 600 Baud rate 1200 Baud rate 2400 Baud rate 4800 Baud rate 9600 8 bits, no parity 7 bits, even parity 7 bits, odd parity Standard printer setting Not documented Data output format (not available in models CFS 300-3 CFS 3K-5 CFS 50K-3) For printout sample see chapter 17.2 Input/display user identification number, max. 6 digits Input/display scale identification number max. 6 digits Details s.Chap.
12.2.2 Models CFS 3K-5, CFS 300-3 Menu block Main menu Menu item Submenu Available Settings Explanation Signal tone switched off Signal tone on, if weighing value within tolerance limits Signal tone on, if weighing value outside tolerance limits Display background illumination off resp.
Baud rate 600 Baud rate 1200 Baud rate 2400 Baud rate 4800 Baud rate 9600 8 bits, no parity 7 bits, even parity 7 bits, odd parity Standard printer setting Not documented Data output format (not available in models CFS 300-3 CFS 3K-5 CFS 50K-3) For printout sample see chapter 17.2 Input/display user identification number, max. 6 digits Input/display scale identification number max. 6 digits on Automatic reference optimization on/off off on Signal tone when key is pressed on / off off Details s.Chap.
13 Configuration of bulk scale Changes may only be carried out by trained specialized personnel. Ex factory, the scales KERN CFS respectively counting systems KERN CCS have been programmed such that no changes are generally necessary. But if there are special conditions of use or if as bulk material balance an other weighing bridge (not preconfigured by KERN) is connected, in the menu block " " the required settings can be made. Technical data Supply voltage: 5 VDC Max.
Menu settings: „F1 oFF“ Call up menu Switch-on balance and during the selftest press The first menu block is displayed. . Press repeatedly or (in CFS 50K-3 models) press again until is displayed. „Pin“ Acknowledge with appears. . The request to enter the password Or as standard password enter four times zero "0000" or the stored password (input see parameter ). (Emergency password „9999“) Acknowledge by „tECH“ „Pin“ „----„ .
(1) Adjusting the bulk scales, all models except for CFS 50K-3 „Cnt“ 1. Internal resolution Press , the internal resolution will be shown. Return to menu by Use . to select the next menu item „Cap“. 2. Position decimal point / capacity „CAP“ Press on display „CAP“; the currently adjusted position of the decimal point will be displayed. Select desired setting with and acknowledge by The currently set capacity is displayed. .
5. Zero adjustment range „0 AUto“ Range of loads in which the display will be set to zero after activation. Press shown. on display „0 AUto“; the current adjustment will be Select desired setting by and confirm with balance returns to the menu. Select the next menu point „0 manl“ with Adjustments are only possible on the reference scale. , the . 6. Manual zero tracking Range of loads in which the display will be set to zero after actuation of the zero key.
(2) Adjustment of bulk scales, model CFS 50K-3 1. Internal resolution Press „Cnt“ , the internal resolution will be shown. Return to menu by . Select the next menu point „dESC“ with . 2. Position decimal point / „dESC“ Press on display „dESC“; the currently adjusted position of the decimal point will be shown. Select desired setting with Use and acknowledge by „dESC“ „0.00“ . CAP to select the next menu item [CAP]. 3.
5. Automatic zero tracking „AZt“ on display change Press , the current setting will be displayed. „AZn“ „2d“ Select desired setting by and confirm with balance returns to the menu. Select the next menu point „0 AUto“ with , the . „AZt“ 6. Manual zero tracking Range of loads in which the display will be set to zero after actuation of the zero key. Press „0 mAnL“ , the current setting will be displayed.
14 Adjustment • • • Provide required adjustment weight, cf. chapter 1. The adjustment weight to be used depends on the capacity of the weight / counting system. Implement adjustment as close to the maximum load as possible. Info about test weights can be found on the Internet at: http://www.kern-sohn.com Observe stable environmental conditions. A warming up time (see chapter 1) is required for stabilization.
Ensure that there are no objects on the weighing pan. Wait for stability display (indicator [] atop . expires), then press Carefully place required adjustment weight into the center of the weighing plate on display „LoAd“. Wait for stability display, then press . After the adjustment the balance will carry out a self-test. Remove adjusting weight during selftest, balance will return into weighing mode automatically.
Ensure that there are no objects on the weighing pan. Wait for stability display (indicator [] atop then press appears), . The currently adjusted adjustment weight (e.g. 6 kg) will be shown. If applicable, change the indicated weight value with the numeric keys. Acknowledge with Display examples model CFS 6K0.1 . Carefully place the indicated weight value for adjustment into the center of the weighing plate on display „LoAd“. Wait for stability display, then press .
14.3 Model KERN CFS 50K-3 Operation Display Switch-on balance and during the selftest press „Pin“ . Use the numeric keys to enter password: Either input four times zero „0000“ as a standard password, or input the user-defined password (input cf. parameter „Pin“ chapter 13). Confirm input by . Select bulk material or reference balance via appeared [] displays the current balance. .
15 Linearization Linearity shows the greatest deviation of a weight display on the scale to the value of the respective test weight according to plus and minus over the entire weighing range. If linearity deviation is discovered during a monitoring of test resources, you can improve this by means of linearization. • • • • • Carrying out linearization is restricted to specialist staff possessing well acquainted with the workings of weighing scales.
Table 2: Required adjustment weights for interconnected bulk scales 1.
15.1 Models CFS 300-3, CFS 3K-5 Operation Display Switch-on balance and during the selftest press . „Pin“ Use the numeric keys to enter password „9999“: Confirm input by „Pin“ „----„ . Select bulk material or reference balance via The appeared [] displays the current balance. . When using as counting system, the bulk material balance as well as the reference balance must be linearized. Both scales must be linearized.
Carefully place fourth adjustment weight into the center of the weighing plate on display „LoAd 4“. Wait for stability display, then press . Ensure that no objects are on the weighing plate on display „LoAd 0“. Wait for stability display, then press . Carefully place fourth adjustment weight into the center of the weighing plate once more on display „LoAd 4“. Wait for stability display, then press .
Use the numeric keys to input password „9999“: Confirm input by „Pin“ „----„ . Select bulk material or reference balance via The appeared [] displays the current balance. . „tECH“ „LoCAL“ When using as counting system, the bulk material balance as well as the reference balance must be adjusted. The adjustment process must be carried out on both balances. Acknowledge with „tECH“ „rEmotE“ . Using select the weighing unit [kg or lb] which will be used for linearization.
16 Second balance interface When using as counting system, the platform must be connected with a suitable cable via the second-balance interface. All models except for CFS 50K-3: 9 pin d-subminiature bushing of the balance Pin no.: Balance connection Pin 1 or 2 EXC+ (5V) Pin 4 or 5 EXC- (0) Pin 7 SIG– Pin 8 SIG+ Interconnection of platform KERN KFP see labelling of load cell Model CFS 50K-3: Pin no.
17 RS 232C interface The balance is typically equipped with a RS 232C interface. The weighing data can be output depending on the setting in the menu either automatically or by pressing or (in CFS 50K-3) via the interface. This data exchange is asynchronous using ASCII - Code. The following conditions must be met to provide successful communication between the weighing balance and the printer. • Use a suitable cable to connect the weighing balance to the interface of the printer.
17.2 Printer operation 17.2.1 Printout example KERN YKB-01N / model CFS 300-3 Counting S1 ID: N Active balance, see chap. 7.3 123456 250.001 g 1.17647 g/pcs 212 pcs User identification number (cf. chapter 12.2) Net weight Average piece weight Quantity 17.2.2 Printout examples KERN YKB-01N / model CFS 3K-5 Counting S1 ID: N 72 Active balance, see chap. 7.3 123456 1.20005 kg 2.49991 g/pcs 480 pcs User identification number (cf. chapter 12.
Totalization 1. Weighing S1 ID: 123456 ABCDEF N 1.15014 kg 2.00011 g/Pcs 575 Pcs Active balance, see chap. 7.3 User identification number (cf. chapter 12.2) Article name (cf. chapter 11) Placed net weight Average piece weight Quantity placed on balance C No. C C 1 1.15014 kg 575 pcs 2. Weighing S1 ID: 123456 ABCDEF N 0.90001 kg 2.00011 g/Pcs 450 Pcs Number weighing processes Total weight Total number of pieces Active balance, see chap. 7.3 User identification number (cf. chapter 12.
17.2.3 Printout examples KERN YKB-01N / CFS 6K0.1, CFS 15K0.2, CFS 30K0.5 Summing up /menu adjustment „F2 PrtForm 1 (cf. chapter12.2) 1. Weighing S1 ID: 123456 ABCDEF N 5.0002 kg 10 g/Pcs 500 Pcs Active balance, see chap. 7.3 User identification number (cf. chapter 12.2) Article name (cf. chapter 11) Placed net weight Average piece weight Quantity placed on balance C No. C C 1 5.0002 kg 500 pcs 2. Weighing S1 ID: 123456 ABCDEF N 2.
Summing up /menu adjustment „F2 PrtForm 2 (cf. chapter12.2) 1. Weighing S1 ID: 123456 ABCDEF N 2.5003 kg G 3.0000 kg T 0.4997 kg 10 g/Pcs 250 Pcs Active balance, see chap. 7.3 User identification number (cf. chapter 12.2) Article name (cf. chapter 11) Placed net weight Gross weight put on Tare weight Average piece weight Quantity placed on balance C No. C C 1 2.5003 kg 250 pcs 2. Weighing S1 ID: 123456 ABCDEF N 5.5003 kg G 6.0000 kg T 0.
Summing up /menu adjustment „F2 PrtForm 3 (cf. chapter12.2) 1. Weighing S1 ID: 123456 ABCDEF N 2.5002 kg G 2.9999 kg T 0.4997 kg 10 g/Pcs 250 Pcs HI 100 PCS LO 90 PCS ---------HI---------- Active balance, see chap. 7.3 User identification number (cf. chapter 12.2) Article name (cf. chapter 11) Placed net weight Gross weight put on Tare weight Average piece weight Quantity placed on balance Upper tolerance limit, see chap. 9.2 Lower tolerance limit, see chap. 9.
3. Weighing S1 ID: 123456 ABCDEF N 1.0002 kg G 1.4999 kg T 0.4997 kg 10 g/Pcs 100 Pcs HI 100 PCS LO 90 PCS ---------OK---------- Active balance, see chap. 7.3 User identification number (cf. chapter 12.2) Article name (cf. chapter 11) Placed net weight Gross weight put on Tare weight Average piece weight Quantity placed on balance Upper tolerance limit, see chap. 9.2 Lower tolerance limit, see chap. 9.2 Target number of pieces within the given tolerance C No. C C 3 4.
17.2.4 Printout examples KERN YKB-01N / model CFS 50K-3 Totalization 1. Weighing LOCAL SCALE ID: 123456 ABCDEFGHIJKL 6.500 kg NET 100 g U. W. 65 PCS Active balance, see chap. 7.3 User identification number (cf. chapter 12.2) Article name (cf. chapter 11) Placed net weight Average piece weight Quantity placed on balance TOTAL 6.500 kg NET 65 TPC 1 NO 2. Weighing LOCAL SCALE ID: 123456 ABCDEFGHIJKL 14.502 kg NET 100 g U. W.
17.3 Remote control instructions 17.3.1 All models except for KERN CFS 50K-3 All inputs finish with (carriage return / line feed). At wrong inputs the command will be preceded by "ER", e.g. order "NN", error message "ER NN". 17.3.2 Control commands PLUxx Call article from data storage T Tare placed weighing vessel T123.456 Use the numeric keys to input tare value e.g. 123.
17.3.4 Model CFS 50K-3 Do not finish inputs with (carriage return / line feed). T Taring Z P Zeroing Printing 17.3.5 User identification, scale identification, store article name SUID xxxxxx User identification number no more than 6 characters SSID xxxxxx Scale identification number no more than 6 characters SSID xx, Storage place 2 characters + comma xxxxxxxxxxxx Article name No more than 12 characters Unavailable in model CFS 50K-3 17.3.
17.4 I/O-function RS-232 Fig.: 9 pin d-subminiature bushing Pin 2 Pin 3 Pin 4 Pin 5 Pin 1 Pin 5 Pin 6 Pin 7 Pin 8 Pin 9 RS232 Shift point RXD TXD VCC GND VB GND OK LOW HI BEEP 5V Connection example with lights CFS-A03 Signal light J2 AC power beep yellow red green COM VB IN1 IN2 IN3 IN4 GND COM OUT1 OUT2 OUT3 OUT4 J1 220VA UOH High-Level Output Voltage UOL Low-Level Output Voltage CFS/CCS-BA-e-1422 2.4 V 0.
18 Servicing, maintenance, disposal Before any maintenance, cleaning and repair work disconnect the appliance from the operating voltage. 18.1 Cleaning Please do not use aggressive cleaning agents (solvents or similar agents), but a cloth dampened with mild soap suds. Ensure that no liquid penetrates into the device. Polish with a dry soft cloth. Loose residue sample/powder can be removed carefully with a brush or manual vacuum cleaner. Spilled weighing goods must be removed immediately. 18.
19 Instant help In case of an error in the program process, briefly turn off the balance and disconnect from power supply. The weighing process must then be restarted from the beginning. Fault Possible cause The displayed weight does not glow. • The balance is not switched on. • The mains supply connection has been interrupted (mains cable not plugged in/faulty). • Power supply interrupted. • Draught/air movement • Table/floor vibrations • Weighing pan has contact with other objects.
19.
20 Declaration of conformity KERN & Sohn GmbH D-72322 Balingen-Frommern Postbox 4052 E-Mail: info@kern-sohn.de Phone: 0049-[0]7433- 9933-0 Fax: 0049-[0]7433-9933-149 Internet: www.kern-sohn.