User guide

19.1 Error messages
Error message Description Possible reasons/
solution
Err 4
Zeroing range exceeded
due to switching-on balance
or pressing (normally
4% max)
Object on the weighing
plate
Overload when zeroing
Inappropriate
adjustment
Damaged weighing cell
Damaged electronics
Err 5
Keyboard error
Unpurposeful
operation of the scale
Err 6
Value outside the A/D
changer range
Weighing plate not
installed
Damaged weighing cell
Damaged electronics
Err 19
Zero point displaced
Remedy:
Adjust / linearize
FAIL H / FAIL L
Adjustment error
Inappropriate
adjustment
Should other error messages occur, switch balance off and then on again. If the
error message remains inform manufacturer.
84 CFS/CCS-BA-e-1422