Procedure when adjusting:
Observe stable environmental conditions.
A warming up time (see chapter 1; Technical Data) is required for stabilization.
Ensure that there are no objects on the weighing plate.
Models FOB-NS:
Balance display
Operation
(example)
Switch off the balance
Press and hold , press 3x at the same
time, release - an internal value is displayed.
(example)
Press , "0" is displayed
Put adjustment weight in the centre of the weighing
plate and press , "CAL" is displayed, followed
by the value of the adjustment weight.
Take away adjustment weight
The balance changes into weighing mode.
Adjustment has now been completed successfully.