Owner manual
KFS-T-BA_IA-e-1213 41
PC
D-SUB9P
Balance
D-SUB9P
9 RS 232C interface
The RS 232C interface allows a bi-directional data exchange from the balance to
external devices. This data exchange is asynchronous using ASCII - Code.
The following conditions must be met to provide successful communication between
the weighing system and the printer.
• Use a suitable cable to connect the display unit to the interface of the printer.
Faultless operation requires an adequate KERN interface cable.
• Communication parameters (Baud rate, parity, see chap. 8) of display device
and printer must match.
Pin allocation of balance output plug:
Pin nr.
Signal
Input/Output
Function
2 RXD Input Receive data
3
TXD
Output
Transmit data
4 DTR Output HIGH
5 GND - Signal ground
6 - -
7 - -
8 - -
9 GND - Signal ground
Interface cable:
TXD
RX
GN
RX
TXD
GN
3
2
5
2
3
5