KERN & Sohn GmbH Ziegelei 1 D-72336 Balingen E-Mail: info@kern-sohn.com Tel: +49-[0]7433- 9933-0 Fax: +49-[0]7433-9933-149 Internet: www.kern-sohn.com Installation instructions platform KERN KXP V20 Version 1.
GB KERN KXP V20 Version 1.0 09/2013 Operating Instruction Platforms Tabel of contents 1 General ........................................................................................................... 3 2 Safety precautions ......................................................................................... 3 3 Setting up the weighing platform ................................................................. 4 3.1 Selecting the site of installation...........................................
1 General These installation instructions cover all information required for the installation and start-up of the following platforms: KXP 6V20LM KXP 15V20M / KXP15V20LM KXP 30V20M / KXP30V20LM KXP 60V20M / KXP 60V20LM KXP150V20M / KXP 150V20LM KXP300V20M 2 Safety precautions Product safety plays an important role at KERN & Sohn. Non-observance of the following instructions can lead to damage to the weighing platform and/or injuries. Before using the weighing platform read these instructions.
3 Setting up the weighing platform 3.1 Selecting the site of installation • • 3.2 3.3 The surface must be able to bear the weighing platform under maximum load at the points of support. At the same time it should be so stable that no vibrations arise during weighing. This is also to be observed when installing the weighing platform in conveyor and similar systems. If possible, vibrations from neighboring machines should not occur at the site of installation.
Packaging / return transport Keep all parts of the original packaging for a possibly required return. Only use original packaging for returning. Reattach supplied transport securing devices. Secure all parts against shifting and damage. 3.4 Ambient conditions Do not use the weighing platform in wet or corrosive environments. Never immerse electronic products into liquids.
3.6 Connecting to the weighing terminal Deadweight cell output KERN KXP V20 weighing platform connection EXC+ See marking of the deadweight cell EXCSIGSIG+ 4 Operating limits The weighing platform is designed so robustly that an occasional exceeding of the maximum weighing load does not lead to damage. The static bearing capacity, i.e. the maximum permissible load, depends on the type of load carrying (position A – C). The maximum static bearing capacity may not be exceeded.
5 Cleaning Clean the weighing platform with a soft cloth soaked with a mild cleaning agent. Take off the load panel and remove any dirt and foreign substances which may have collected underneath it. Do not use any hard objects to do so. Do not open the weighing platform. 6 Technical data 6.
6.2 Technical data of the weighing cell Sensitivity 2.0 ±0.
6.3 Preload, Deadload and Overload settings Kern model max. Preload* (kg) * = additional initial load KXP 6V20 LM KXP 15V20 M KXP 15V20 LM KXP 30V20 M KXP 30V20 LM KXP 60V20 M KXP 60V20 LM KXP 150V20 M KXP 150V20 LM KXP 300V20 M 9 Center Overload Protection circa (kg) Corner Overload Protection circa (kg) 2.14 2.14 4.48 4.48 9.02 4.48 9.02 9.02 13.86 13.86 8.5 23 23 46 46 85 85 200 270 550 4.8 12 12 24 24 48 48 120 120 240 1.86 2.86 2.86 10.52 10.52 35.52 35.52 90.98 136.14 186.