Manual

KEN-TM / KET_TM-BA-e-1420 169
16.3.3 Retrieve tare value
Command: OT CR LF
Answer:
1 2 3 4 - 12 13 14 - 16 17 18 19
O T
˽
Tare
˽
Unit
˽
CR LF
Tare value: - 9 characters with right-justified alignment
Unit: - 3 characters with left-justified alignment
Tare values are always issued in adjusting unit
16.3.4 Set tare value
Command: UT˽TARA CR LF, (TARA = tare value)
Possible responses:
UT˽OK CR LF
Command executed
UT˽I CR LF
Command valid but can currently not be executed
ES CR LF
Invalid entry
Set tare value:
Mark decimal places with a dot
Without weighing unit
16.3.5 Send stable weighing value in standard weighing unit
Command: S CR LF
Possible responses:
S˽A CR LF
S˽E CR LF
Command accepted; executing command
Time limit for stabilisation of weighing scale display was exceeded
S˽I CR LF
Command valid but can currently not be executed
1
2-3
4
5
6
7-15
16
17-19
20
21
S
˽
Stability
icon*
˽
Signs Weight
˽
Unit CR LF