Manual
KEN-TM / KET_TM-BA-e-1420 171
16.3.7 Send weighing values for all platforms in default weighing unit
immediately
Command: SIA CR LF
Possible responses
SIA
˽
I CR LF
Command valid but can currently not be executed
1
2
3
4
5
6
7-15
16
17-19
20
21
S I
˽
Stability
icon
˽
Signs Weight
˽
Unit CR LF
n - Number of platform
Weight - 9 characters with right-justified alignment
Unit - 3 characters with left-justified alignment
Example (two connected platforms):
S I A CR LF
For sent command see chap. 16.1
P 1˽?˽˽˽˽˽˽1 1 8 . 5˽g˽˽CR LF
Command executed,
Weighing values of platforms will be issued in default
weighing unit
P 2˽˽˽˽˽˽˽˽˽3 6 . 2˽k g˽CR LF
16.3.8 Send stable weighing value in current weighing value
Command: SU CR LF
Possible responses:
SU
˽
A CR LF
SU˽E CR LF
Command accepted; executing command
Time limit for stabilisation of weighing scale display was
exceeded
SU˽I CR LF
Command valid but can currently not be executed
SU˽A CR LF
Command accepted; executing command
1
2
3
4
5
6
7-15
16
17-19
20
21
S U
˽
Stability
icon
˽
Signs Weight
˽
Unit CR LF
Example:
S U CR LF
For sent command see chap. 16.1
S U _ A CR LF
Command accepted; executing command
S U˽˽˽-˽˽1 7 2 . 1 3 5˽N˽˽CR LF
Command executed. Weighing value will issued in current
weighing unit