KEN-TM / KET_TM-BA-e-1420 173
16.3.12 Start continuous output in current weighing unit
Command: CU1 CR LF
Possible responses:
CU1˽I CR LF
Command valid but can currently not be executed
CU1˽A CR LF
Command accepted; executing command
Weighing values will be issued in current weighing unit
1
2
3
4
5
6
7-15
16
17-19
20
21
S U I
Stability
icon
˽
Signs Weight
˽
Unit CR LF
16.3.13 Stop continuous output in current weighing unit
Command: CU0 CR LF
Possible responses:
CU0˽I CR LF
Command valid but can currently not be executed
CU0˽A CR LF
Command accepted; executing command
16.3.14 Set value for “Lower Threshold“
Message: DH˽XXXXX CR LF (XXXXX = threshold value)
Possible responses:
DH˽OK CR LF
Command executed
ES CR LF
Command not accepted
16.3.15 Set value for “Upper Threshold“
Command: UH˽XXXXX CR LF (XXXXX = threshold value)
Possible responses:
UH˽OK CR LF
Command executed
ES CR LF
Command not accepted