Instruction Manual
KFB/KFN-TM-BA_IA-e-1323 49
Symbols:
ST
Stable value
US
Instable value
GS / GW
Gross weight
NT
Net weight
TW
Tare weight
NO
Number weighing processes
TOTAL
Total of all individual weighings
<lf>
Space line
<lf>
Space line
• Counting
10.3 Output log (continuous output)
• Weighing
HEADER1: ST=STABLE, US=UNSTABLE
HEADER2: NT=NET, GS=GROSS
10.4 Remote control instructions
Command Function Printout examples
S Stable weighing value for the weight is
sent via the RS232 interface
ST,GS 1.000KG
W Weighing value for the weight (stable or
unstable) is sent via the RS232 interface
US,GS 1.342KG
ST,GS 1.000KG
T No data are sent, the balance carries out
the tare function.
-
Z No data are sent, the zero-display
appears.
-
P Quantity will be sent via the RS232-
interface
10PCS
****************************
PCS 100
****************************