Instruction Manual

KXE-TM-BA_IA-e-1411 35
14 Technician menu
In verified weighing systems the access to the technician menu is
locked.
In order to unlock the access, the seal must be destroyed and the
adjustment switch be actuated (see chap. 6.8).
Attention:
After destruction of the seal the weighing system must be re-verified by
an authorised agency and a new verification wire/seal mark fitted before
it can be reused for applications subject to verification.
Call up menu:
Actuate adjustment switch, „SET followed by01 CSP“ is displayed.
Use and to switch over between the following menu items:
Configure
Adjustment
Enter password
Change settings of the unit
To exit the menu / back to weighing mode, actuate adjustment switch.