Instruction Manual
KXE-TM-BA_IA-e-1411 37
Configuration menu overview:
Factory settings or allowed settings for verified weighing systems are marked with *.
No. Function
Available settings
Paramete
rs
Description
CSP-00 Position decimal
point
d 0 No decimal place
d 0.0 1 decimal place
d 0.00* 2 decimal places
d 0 3 decimal places
d 0.0000 4 decimal places
CSP-01 Balance capacity
[Max]
999999
000000
The possible capacities and
factory settings depend on the
connected weighing bridge.
CSP-02 Readability [d] /
verification value [e]
1. weighing range
1 The possible resolutions and
factory settings depend on the
connected weighing bridge.
2
5
10
20
50
CSP-03 Readability [d] /
verification value [e]
2. weighing range
1 The possible resolutions and
factory settings depend on the
connected weighing bridge.
2
5
10
20
50