Instruction Manual

KXE-TM-BA_IA-e-1411 39
14.2 Adjust / Menu item
The weight to be used depends on the capacity of the scale. Carry out
adjustment as near as possible to the scale’s maximum weight. Info about
test weights can be found on the Internet at: http://www.kern-sohn.com.
Observe stable environmental conditions. Stabilisation requires a certain
warm-up time.
Actuate adjustment switch, „SET“ followed by „01 CSP“ is displayed.
Invoke menu item using or and confirm by .
The adjustment of the zero point is displayed.