User guide

Use the navigation buttons to select desired settings
Example unit 20:
Number of balancing units "2 (a), standard weighing unit „kg(b),
Confirm with TARE key, the current standard weighing unit is indicated for 2
seconds (e.g. „kg) followed by the display „000001 for adjustment to the
capacity (max.).
Settings maximum:
Description
(c)
Standard weighing unit
(d-h)
Capacity (max)
(i-n)
Weight value of adjusting weight
With the navigation keys (see chapter 2.1.1) enter the desired capacity.
Example 015000
15.000kg
Confirm with TARE key, the display "000002" appears for the input of the weight
value for adjusting weight.
Enter the desired weight value with the navigation keys.
88 KXS/KXG-TM-BA_IA-e-1420