User guide

Example 4: 5-point linearization/ 30 kg (5kg, 5kg, 10 kg, 10kg)
Display
Button
Explication
03 CLn TARE Linearization function
L0 TARE zero point
L11 TARE First adjustment weight (5 kg)
L21 TARE Second adjustment weight (5kg), factor 1, i.e., the same size as the first
L32 TARE Third adjustment weight (10 kg); factor 2, i.e. twice as high as the first
L42 Fourth adjustment weight (10 kg); factor 2, i.e. twice as high as the first
L51 BG/NEt Terminate linearization (5-point linearization)
03 CLn
Example 4: 7-Point linearization / 30 kg (1 kg, 2 kg, 5 kg, 10 kg, 2 kg, 10 kg)
Display
Button
Explication
03 CLn TARE Linearization function
L0 TARE zero point
L11 TARE First adjustment weight (1 kg)
L22 TARE Second adjustment weight (2 kg); factor 2, i.e. twice as high as the first
L3 5 TARE Third adjustment weight (5 kg); Factor 5 i.e. 5 times as high as the first
L4 A TARE Forth adjustment weight (10 kg); Factor A i.e. 10 times as high as the first
L52 TARE Fifth adjustment weight (5kg); factor 2, i.e. twice as high as the first
L6 A TARE Sixth adjustment weight (10 kg); Factor A i.e. 10 times as high as the first
L71 BG/NEt Terminate linearization (7-point linearization)
03 CLn
96 KXS/KXG-TM-BA_IA-e-1420