9  Using scale 
9.1  Weighing 
  Switch on the scale with the [ON/OFF] key. The diagnostic scale self-check is 
performed and then the software version is displayed. The scale is ready for 
weighing when the „0,00 kg” weight display is shown. 
Direction: The [→0←] key makes it possible to zero the scale if necessary and 
at any time.  
  Place a person in the middle of the scale. Wait until the stability display (ο) is 
shown and then read the weighing result. 
Direction: 
If a person is heavier than the weighing range, the display will show the „Err” symbol 
(= overload).  
9.1.1  Weighing with MWS 
Due to the great dimensions and the big weighing range especially suitable for 
weighing of immobile patients on transportation stretchers, wheelchairs or overweight 
patients in the adiposity range 
9.1.1.1  Weighing with transportation stretcher or wheelchair 
 Place transportation stretcher/wheelchair with patient on the centre of the 
scales 
 Arrest the brakes of the transportation stretcher/of the wheelchair 
Do not leave the patient unattended 
 When the patient is lying/sitting quietly, read weighing value 1 
 Loosen the brakes and carefully pull off the transportation stretcher/wheelchair 
with patient 
 After that weigh the transportation stretcher/wheelchair without patient and 
subduct this weight from weighing value 1, from there results the patient’s 
weight. 
        MPS / MTS / MXS / MWS-BA-e-1417 44 