Manual

Sauter GmbH
Tieringerstr. 11-15
D-72336 Balingen
E-Mail: info@sauter.eu
Tel: +49-[0]7433- 9976-174
Fax: +49-[0]7433-9976-285
Internet: www. sauter.eu
Instruction Manual
SD
SD-BA-e-1112 1
Model: SD, Spring Tester
Thank you for buying a SAUTER Spring Tester. We hope
you are pleased with your high quality force gauge with its
big functional range. If you have any queries, wishes or
helpful suggestions, do not hesitate to call our service
number.
1.
Working Conditions
10°C to 30°C / 15% up to 80% humidity
2.
Operation
a. Display
(1) Measuring Result
(2) Measuring Units
(3) Activation of PRINT Function
(4) Indication of power charging status
(5) PEAK or AUTO-PEAK Mode
(6) Average value of stored peak values
(7) Force direction
(8) Occupancy of storing spaces
(9) AVERAGE- or Saving Mode
b. Operating keys
ON / OFF:
ON / OFF key
(For ON, press 1 sec.)
UNIT:
- Press shortly: Select unit:
N, kg or lb
- Press for 2 sec.: Display return
ZERO:
Three functions:
- Zeros the measuring result (Tara function)
- Cleans the peak value (in Peak mode)
- Saves a setting (in SET mode)
SET:
- 1 x Press: Upper Limit [HidT / Hi Lt]. To
change press: ▲ or ▼ (see section c.)
- 1 x Press: Lower Limit [LodT]. To change
press ▲ or ▼ (see section c.)
- 1 x Press: Minimum limit to save peak
values in the instrument’s memory.
[LE.SET / Lo Pe]. Please see section „Min
Limit Peak Save“. To change press: or
(Only in „Peak-Mode“ active) (see
section g.)
- 1 x Press: Auto-Off Function. Turns the
instrument off after a here defined time
period in sec. [P.OFF]. To change press: ▲
or ▼ (Only in „Battery-Mode“ active)
- 1 x Press: Peak-Freeze-Time [PE.2E /
A.PE / HoldT]: Time period in which a peak
value is being shown in the display in sec.
To change press: ▲ or ▼
- To the printer (Print)
1 x Press: Saves the Settings
Back Light:
Hi Lt
Lo dT
Lo Pe
P.OFF
HoldT

Summary of content (4 pages)