User guide

3-6 DP-V1 Service
3-38
- PROFIBUS DP Compatible Network Unit DL-PD1 User's Manual (IL) -
Executing Communication
3
*12 Of the motion commands "Tolerance tuning request", "Two-point tuning HIGH side
2nd point confirmation", "Two-point tuning LOW side 2nd point confirmation", "One-
point tuning request for diff. count filter", "Two-point tuning 2nd point confirmation for
diff. count filter", reads the execution result for the item last requested.
*13 This parameter works if "Bank switching method" is set to "button."
*14 These parameters are OFF only if data "0" is written and the wiring is configured to
turn off external input.
*15 When using IL-300/IL-600, -999.99 to +999.99.
When using IL-2000, -9999.9 to +9999.9.
*16 A write control error results if data is rewritten to an expansion unit.
*17 To have the sensor amplifier reflect the settings written in data numbers nn92 to
nn95, it is necessary to set data number nn91 to "1" (user setting), or to set the
external input to "user setting" via button operation on the sensor amplifier.
*18 An attempt to write this data results in an error. (Reading is results in "0.")
*19 The read value of the two-digit value can be translated to a binary number, so that the ON/
OFF state of each bit can be noted to know the status of judgment output and alarm output.
Assume the read data is "05":
Translating "05" to a binary number gives "0101".
N.O. mode
*20 The external input status is "1" when the external input is ON, and is "0" when the
external input is OFF. Bits 0, 1, 2, and 3 correspond to external inputs 1, 2, 3, and 4,
respectively.
Assume the read data is "06":
Translating "6" to a binary number gives "0110".
*21 Stores the read string in the remote registers in ASCII code.
For "IL-1": RWr03=312Dh, RWr02=4C49h
Bit
Judgment output
N.O. mode N.C. mode
0
0: HIGH judgment output OFF
1: HIGH judgment output ON
0: HIGH judgment output ON
1: HIGH judgment output OFF
1
0: LOW judgment output OFF
1: LOW judgment output ON
0: LOW judgment output ON
1: LOW judgment output OFF
2
0: GO judgment output OFF
1: GO judgment output ON
0: GO judgment output ON
1: GO judgment output OFF
3
0: Alarm output OFF
1: Alarm output ON
0: Alarm output ON
1: Alarm output OFF
Reference
0101
Bit 3: Alarm output OFF
Bit 2: GO judgment output ON
Bit 0: HIGH judgment output ON
Bit 1: LOW judgment output OFF
Reference
0110
External input 4: OFF
External input 3: ON
External input 1: OFF
External input 2: ON