Owner's manual
13
- PROFINET Communication Unit DL-PN1 User's Manual (IG) -
90 External input response 5
When the request reception
to the "External input request
5" is complete, the
corresponding bit is set.
*1
2byte
WORD
R
bit0 : Reserved for system
bit1 : Slot 1
:
bit15 : Slot 15
91 to 95 Reserved for system
96
1st Sensor Head 1
Identification
Stores the type of the slot
number 1 sensor head (light
transmitter).
4byte
DINT
R
1:
Light transmitter of IG-028
2:
Light transmitter of IG-010
3: Not recognized or
connected
:: : :: :
99
4th Sensor Head 1
Identification
Stores the type of the slot
number 04 sensor head (light
transmitter).
4byte
DINT
R
1:
Light transmitter of IG-028
2:
Light transmitter of IG-010
3: Not recognized or
connected
100 to
111
Reserved for system
112
1st Sensor Head 2
Identification
Stores the type of the slot
number 1 sensor head (light
receiver).
4byte
DINT
R
1: Light receiver of IG-028
2: Light receiver of IG-010
3: Not recognized or
connected
:: : :: :
115
4th Sensor Head 2
Identification
Stores the type of the slot
number 04 sensor head (light
receiver).
4byte
DINT
R
1: Light receiver of IG-028
2: Light receiver of IG-010
3: Not recognized or
connected
116 to
127
Reserved for system
128 Product code 1
Stores the product code of the
slot number 01 sensor amplifier.
2byte
WORD
R
:: : ::
131 Product code 4
Stores the product code of the
slot number 04 sensor amplifier.
2byte
WORD
R
132 to
143
Reserved for system
144 Revision 1
Stores the revision of the slot
number 01 sensor amplifier.
2byte
WORD
R 0101H to FFFFH
:: : :: :
147 Revision 4
Stores the revision of the slot
number 04 sensor amplifier.
2byte
WORD
R 0101H to FFFFH
148 to
159
Reserved for system
160
Comparator value 0 (Slot
Number 1)
Used when reading the
comparator value (P.V. value)
of slot number 01.
*2
4byte
DINT
R
-99.999 to +99.999:: : ::
163
Comparator value 0 (Slot
Number 4)
Used when reading the
comparator value (P.V. value)
of slot number 04.
*2
4byte
DINT
R
164 to
175
Reserved for system
176
Number of decimal places
(Current value 0, Slot
Number 1)
Used when reading the
decimal position of the slot
number 01 comparator value
(P.V. value).
4byte
DINT
R
:: : ::
Index
(DEC)
Function name Description
Data
type
Attribute
Value