4-2 Basic Settings and Advanced Settings
4-17
IG-E
4
Setting Various Functions
This is displayed when any of "Edge control/Positioning mode", "Glass edge mode" or "Pin
position measurement mode" is set for "1. Measurement mode" (page 4-8). (If
measurement modes other than the above are set, this screen is skipped.)
For details, refer to the "1. Measurement mode" (page 4-8).
The response time is the time from when the sensor head starts the measuring operation
and the value at the measurement position is established as an internal measurement
value (R.V.) and judgment value (P
.V.), to time until when the judgment is output. When the
average number of times is increased, the response time becomes longer, and the internal
measurement value (R.V.) and judgment value (P.V.) stabilize.
*1 When "Pin interval judgment mode" or "Pin diameter judgment mode" is set for "1.
Measurement mode" (page 4-8), "JUR" and "" to "" cannot be set.
Response time list
*1 Response time when modes other than "Pin interval judgment mode" and "Pin
diameter judgment mode" are set for " 1. Measurement mode" (page 4-8)
2. Measurement direction
Item Setting range Default value
Measurement Direction VQR (Top), DVO (Bottom) VQR
3. Average number of times (Response time)
Item Setting range
*1
Default value
Average number of times
JUR / / / / /  /  /  /  /  / 
/  /  / 

Configuration
Average
number of
times
Sampling cycle
(ms)
R.V. update cycle
(ms)
Response time
(ms)
*1
hsp 1 0.49 0.49 1.96
11
0.98
0.98
3.92
2 2 4.90
4 4 6.86
8 8 10.78
16 16 18.62
32 32 34.30
64 64 65.66
128 128 128.38
256 256 253.82
512 512 1.96 505.68
1024 1024 3.92 1009.40
2048 2048 7.84 2016.84
4096 4096 15.68 4031.72
Reference
The same values as the above are assigned for the analog output.