User guide

4-2 Basic Settings and Advanced Settings
4-25
IG-E
4
Setting Various Functions
Auto peak hold
When the internal measurement value (R.V.) goes beyond the specified trigger level, the
sampling begins. When it falls below the trigger level, the sampling finishes. When the
sampling finishes, the internal measurement value (R.V.) maximized during the sampling
period is held as a judgment value (P.V.).
The external input is not required.
*1 "11. External input" (page 4-33)
*2 After the power is turned on or reset is input, the judgment value (P.V.) is [] until
the judgment value (P.V.) is first held.
*3 While the
timing input is ON, the sampling does not start. If the sampling is already in
process, it continues.
*4 The main unit (IG-1000/IG-1500) displays the calculation value (CALC value) when the
calculation function is used.
"4-4 Calculation Function" (page 4-47)
Trigger level setting screen
When the auto peak hold is set, the display goes to the trigger level setting screen.
Press S / T button on the trigger level setting screen to specify the trigger level.
Reference
The operation by the timing input is the same regardless of the "6. Timing
input" (page 4-27) setting.
After the sampling starts, if the internal measurement value (R.V.) becomes
"" before the sampling falls below the specified trigger level, the
sampling is automatically finished and the maximized value during the
sampling period is held as a judgment value (P.V.).
Display value
Time
Timing Input
Trigger level
Hysteresis
ON
OFF
Power ON or
Reset input *
1
*
3
*
2
Judgment value (P.V.)
Internal measurement value (R.V.)*
4
Item Setting range Default value
Trigger level -99.999 to 99.999 0.100
Reference
The hysteresis presents the same value as the one set for "8. Hysteresis"
(page 4-30).

LASER
BANK
0
1
2
3
HI
GO
LO
R.V.
ANALOG
HI SHIFT
ZERO SHIFT
TIMING
LO
ALIGNMENT
VT)
HOLD CALC
CHECK
#WVQ2
LASER
BANK
0
1
2
3
HI
LO
R.V.
ANALOG
HI SHIFT
ZERO SHIFT
TIMING
LO
ALIGNMENT
*.&
GO
HOLD CALC
CHECK
QHH
LASER
BANK
0
1
2
3
HI
LO
R.V.
ANALOG
HI SHIFT
ZERO SHIFT
TIMING
LO
ALIGNMENT
&.;
GO
HOLD CALC
CHECK
5. Hold function
Trigger level setting screen
7. Delay timer
[MODE]
or X
[MODE]
or X