User guide
4-4 Calibration Function
4-47
IL-E
4
Setting Various Functions
When using the calculation function
Go to Step 2 of "When calibration function is disabled".
6
Press the [SET] button. (SET2 confirmation operation)
After [SEt2] blinks several times on the sub display (lower
level), [End] blinks.
This completes the calibration.
7
Press the
X
button.
The main screen is restored.
X
button
Main screen
1
Press [MODE] button and X button for approx. 2
seconds on the main screen at the same time.
[CALib] is displayed on the main display (upper level).
2
Press
S
/
T
button to set [
HEAd
] and press
X
button.
Press [MODE] +
X button
2 seconds or more
at the same time
X
button
'P&
LASER
BANK
0
1
2
3
HI
GO
LO
R.V.
ANALOG
HI SHIFT
ZERO SHIFT
TIMING
LO
HOLD CALC
CHECK
Main screen
LASER
BANK
0
1
2
3
HI
GO
LO
R.V.
ANALOG
HI SHIFT
ZERO SHIFT
TIMING
LO
HOLD CALC
CHECK
Setting value
Description
HEAd The internal measurement value (R.V.) calibration is set.
CALc The calculated value (CALC) calibration is set.
*'#&
LASER
BANK
0
1
2
3
HI
GO
LO
R.V.
ANALOG
HI SHIFT
ZERO SHIFT
TIMING
LO
%#.KD
HOLD CALC
CHECK
Calibration setting