User guide
4-4 Calibration Function
4-49
IL-E
4
Setting Various Functions
4
Insert workpiece A for which the measured value is
known, and using the
S
/
T
buttons, set the "CALC
you wish to display (1)" for that workpiece A.
The current internal measurement value is displayed on the
main display (upper level). [SEt1] and the "CALC you wish to
display (1)" are alternately displayed on the sub display (lower
level).
IL-S025/IL-030/IL-S065/IL-065/IL-100
IL-300/IL-600
IL-2000
[SET] button
[SEt1]
5
Press the [SET] button. (SET1 confirmation operation)
After [SEt1] blinks several times on the sub display (lower
level), [SET2] is displayed.
[SET] button
[SEt2]
6
Insert workpiece B for which the measured value is
known, and using the
S
/
T
buttons, set the "CALC
you wish to display (2)" for that workpiece B.
The current internal measurement value is displayed on the
main display (upper level). [SEt2] and the "CALC you wish to
display (2)" are alternately displayed on the sub display (lower
level).
IL-S025/IL-030/IL-S065/IL-065/IL-100
IL-300/IL-600
IL-2000
[SET] button
7542
LASER
BANK
0
1
2
3
HI
GO
LO
R.V.
ANALOG
HI SHIFT
ZERO SHIFT
TIMING
LO
%%
HOLD CALC
CHECK
Item Setting range
Default value
set1 -99.999 to 99.999 5.000
Item Setting range
Default value
set1 -999.99 to 999.99 50.00
Item Setting range
Default value
set1 -9999.9 to 9999.9 500.0
LASER
BANK
0
1
2
3
HI
GO
LO
R.V.
ANALOG
HI SHIFT
ZERO SHIFT
TIMING
LO
HOLD CALC
CHECK
CALC value
Setting you wish to display
LASER
BANK
0
1
2
3
HI
GO
LO
R.V.
ANALOG
HI SHIFT
ZERO SHIFT
TIMING
LO
HOLD CALC
CHECK
CALC value
Setting you wish to display
Item Setting range
Default value
set2 -99.999 to 99.999 10.000
Item Setting range
Default value
set2 -999.99 to 999.99 100.00
Item Setting range
Default value
set2 -9999.9 to 9999.9 1000.0