User guide

4-4 Calibration Function
4-51
IL-E
4
Setting Various Functions
Main screen
1
Press [MODE] button and X button for approx. 2
seconds on the main screen at the same time.
[CALib] is displayed on the main display (upper level).
2
Press
S
/
T
button to set [
CALc
] and press
X
button.
Press [MODE] +
X button
2 seconds or more
at the same time
3
Press
S
/
T
button to set [
USr.3p
] and press
X
button.
[SET] button
[SEt1]
[SET] button
4
Workpiece A is set closer to the head connected to the
main unit allowing for the possible workpiece positional
deviation. Then set the "CALC value you wish to display
(1)" for that workpiece A by S / T buttons.
The current internal measurement value is displayed on the main
display (upper level). [SEt1] and the "CALC you wish to display
(1)" are alternately displayed on the sub display (lower level).
IL-S025/IL-030/IL-S065/IL-065/IL-100
IL-300/IL-600
IL-2000

LASER
BANK
0
1
2
3
HI
GO
LO
R.V.
ANALOG
HI SHIFT
ZERO SHIFT
TIMING
LO

HOLD CALC
CHECK
Setting value
Description
HEAd The internal measurement value (R.V.) calibration is set.
CALc The calculated value (CALC) calibration is set.
%#.E
LASER
BANK
0
1
2
3
HI
GO
LO
R.V.
ANALOG
HI SHIFT
ZERO SHIFT
TIMING
LO
%#.KD
HOLD CALC
CHECK
Calculation calibration setting
7542
LASER
BANK
0
1
2
3
HI
GO
LO
R.V.
ANALOG
HI SHIFT
ZERO SHIFT
TIMING
LO
%%
HOLD CALC
CHECK
Setting value
Description
Default value
dEFLt Standard c
USr.2p User setting (Two-point calibration)
USr.3p User setting (Three-point calibration)

LASER
BANK
0
1
2
3
HI
GO
LO
R.V.
ANALOG
HI SHIFT
ZERO SHIFT
TIMING
LO

HOLD CALC
CHECK
CALC value
Setting you wish to display
Head connected
to main unit
Main unit side
Workpiece A
Head connected to
expansion unit
Item Setting range
Default value
set1 -99.999 to 99.999 5.000
Item Setting range
Default value
set1 -999.99 to 999.99 50.00
Item Setting range
Default value
set1 -9999.9 to 9999.9 500.0