User guide

4-4 Calibration Function
4-53
IL-E
4
Setting Various Functions
[SEt3]
8
Insert workpiece B for which the measured value is
known, and using the
S
/
T
buttons, set the "CALC
you wish to display (3)" for that workpiece B.
The current internal measurement value is displayed on the main
display (upper level). [
SEt3
] and the "CALC you wish to display
(3)" are alternately displayed on the sub display (lower level).
IL-S025/IL-030/IL-S065/IL-065/IL-100
IL-300/IL-600
IL-2000
[SET] button
9
Press the [SET] button. (SET3 confirmation operation)
After [SEt3] blinks several times on the sub display (lower
level), [End] is displayed.
This completes the calibration.
X
button
10
Press the
X
button.
The main screen is restored.

LASER
BANK
0
1
2
3
HI
GO
LO
R.V.
ANALOG
HI SHIFT
ZERO SHIFT
TIMING
LO
HOLD CALC
CHECK

CALC value
Setting you wish to display
Head connected
to main unit
Head connected to
expansion unit
Workpiece B
Item Setting range
Default value
set3 -99.999 to 99.999 10.000
Item Setting range
Default value
set3 -999.99 to 999.99 100.00
Item Setting range
Default value
set3 -9999.9 to 9999.9 1000.0
'P&
LASER
BANK
0
1
2
3
HI
GO
LO
R.V.
ANALOG
HI SHIFT
ZERO SHIFT
TIMING
LO
HOLD CALC
CHECK
Main screen