User guide

3-4 Setting the Tolerance Setting Value
3-17
IL-E
3
Basic Operations
1 point tuning
When there is a measurement object (herein masterwork) that can be understood from the
standard step, based on the step value (P.V.) of the masterwork, 2 times the Judgment
Value HIGH side setting value (upper limit) and 1/2 of the Judgment Value LOW side setting
value (lower limit) can be set automatically.
If the P.V. (Judgment Value) is A, HIGH = |(A) x 2|, and LOW = |(A) ÷ 2| will be the settings.
If the tuning result surpasses the setting range, the setting value will be the setting range
limit value.
1
From the main screen press the W / X button multiple times, from the sub-
display (bottom step) the R.V. (Internal Measurement Value) display
screen will appear.
"Sub Display (Lower Level)" (page 3-4)
2
After the step measurement by the masterwork (after P.V. is fixed), press
and hold the [SET] button.
3
The HIGH side setting value (upper limit value) will be set a 2 times the
value of the P.V. (Judgment Value), and the LOW side setting value (lower
limit value) will be set as 1/2 of the setting value.
(Example) When the P.V. (Judgment Value) is 4000, it will be as follows.
HIGH side setting value: |(4.000) x 2| = 8.000
LOW setting side value: |(4.000) ÷ 2| = 2.000
Point
When the P.V. (Judgment Value) is [-----], the 1 point tuning cannot
occur. If it does occur, [no.vAL] will be displayed on the main screen,
and the number of points will decrease. The setting value cannot be
changed.


R.V (Internal Measurement Value)
P.V (Judgment Value)