User guide

3-18
IL-E
3
Basic Operations
3-5
Zero Shift Function (Shifting the Internal Measurement Value (R.V.))
The internal measurement value (R.V.) is shifted (offset) to an arbitrary shift target value.
The judgment value (P.V.) is shifted (offset) as well.
On a main unit using the calculation function, the calculated value (CALC value) will be
shifted (offset) to the shift target value instead of R.V.
The following two methods can be used.
Press the [ZERO SHIFT] button (within 1 second).
Set the external input (zero shift input) to ON for 20 ms or more.*
* When the zero shift input is set for the external input 1 (pink wire), the zero shift is
enabled when the input is triggered..
"11. External Input" (page 4-30)
1
Press the W / X button several times on the main screen. To display the
shift target value on the sub display (lower level).
"Sub Display (Lower Level)" (page 3-4)
2
Press S / T button to set the shift target value.
IL-S025/IL-030/IL-S065/IL-065/IL-100
IL-300/IL-600
IL-2000
3
Press W / X button to return the sub display to the original screen.
Setting the Shift Target Value
Item Setting range
Default value
Shift target value -99.999 to 99.999 0.000
Item Setting range
Default value
Shift target value -999.99 to 999.99 0.00
Item Setting range
Default value
Shift target value -9999.9 to 9999.9 0.0
GO
HOLD CALC
CHECK

LASER
BANK
0
1
2
3
HI
LO
R.V.
ANALOG
HI SHIFT
ZERO SHIFT
TIMING
LO
ALIGNMENT

Shift target value
[SHIFT]
ON